Спектрометр затухания в полости CO2 с чувствительностью обнаружения на уровне ppb

  • role: First author第一作者
  • Affiliation:

    Anhui Institute of Optics and Fine Mechanics, Hefei Institute of Physical Science, Chinese Academy of Sciences, Hefei 230031, China

    University of Science and Technology of China, Hefei 230026, China

  • Email:yliu@aiofm.ac.cn
  • Introduction:刘 英(1998-),女,四川宜宾人,博士研究生,2018年于安徽大学获得学士学位,主要从事激光吸收光谱方面的研究。E-mail: yliu@aiofm.ac.cn
LIU Ying12,  
  • Affiliation:

    Department of Mechanical and Automation Engineering, The Chinese University of Hong Kong, Hong Kong 999077, China

HU Mai3,  
  • Affiliation:

    Anhui Institute of Optics and Fine Mechanics, Hefei Institute of Physical Science, Chinese Academy of Sciences, Hefei 230031, China

WANG Xingping1,  
  • Affiliation:

    Anhui Institute of Optics and Fine Mechanics, Hefei Institute of Physical Science, Chinese Academy of Sciences, Hefei 230031, China

XU Zhenyu1,  
  • Affiliation:

    Anhui Institute of Optics and Fine Mechanics, Hefei Institute of Physical Science, Chinese Academy of Sciences, Hefei 230031, China

HE Yabai1,  
  • role: Corresponding author通信作者
  • Affiliation:

    Anhui Institute of Optics and Fine Mechanics, Hefei Institute of Physical Science, Chinese Academy of Sciences, Hefei 230031, China

  • Email:kanruifeng@aiofm.ac.cn
  • Introduction:阚瑞峰(1977-),男,辽宁锦州人,博士,研究员,2000年于长春理工大学获得学士学位;2003年、2006年于中国科学院安徽光机所分别获得硕士与博士学位,主要从事光学与光谱学方面的研究。E-mail:kanruifeng@aiofm.ac.cn
KAN Ruifeng1*

реферат

Диоксид углерода (Carbon dioxide, CO2) является главным парниковым газом в атмосфере, характеризующимся высоким фоновым атмосферным уровнем и небольшими годичными изменениями. Поэтому высокоточный мониторинг его концентрации является важным этапом для достижения цели «двойного углеродного следа». В данной статье на основе технологии спектроскопии затухания в оптической полости с непрерывной волной была создана установка для детекции CO2 с чувствительностью до ppb. В системе был выбран центральный длина волны 6251,760 см-1 для поглощательной линии CO2, разработана сверхвысокоточная (более 300 000) кварцевая резонаторная полость Фабри–Перо и высокопроизводительный модуль контроля температуры и давления. Изменения температуры и давления газа в резонаторе за 24 часа составляли менее 0,07 ℃ и 15 Па соответственно. Результаты дисперсии Аллана показали, что система при оптимальном времени интеграции 303 с может достигать предела обнаружения 0,7×10-12 см-1, что соответствует минимальной обнаруживаемой концентрации CO2 в 1,6 ppb. В широком диапазоне концентраций CO2 коэффициент линейной корреляции отклика системы составил 0,99994. В заключение, при времени отклика 10 с система осуществляла непрерывное наблюдение концентрации CO2 в атмосфере в течение 2 дней; результаты были высоко согласованы с данными коммерческого прибора (Picarro, G2401), а относительное отклонение после устранения влияния дыхания человека было менее 6‰. Система отличается простой структурой, низкой стоимостью и чрезвычайно высокой чувствительностью, что делает ее перспективной для мониторинга следовых газов.

ключеви́че слова́

Обнаружение CO2;Оптический датчик;Спектроскопия затухания в полости;Очень высокая чувствительность

References

  1. 1.
    BOUCHER O, BORELLA A, GASSER T, et al. On the contribution of global aviation to the CO2 radiative forcing of climate[J]. Atmospheric Environment, 2021, 267: 118762. doi:
  2. 2.
    LIU W, WEI N N, WANG G H, et al. Quantitative estimation source of urban atmospheric CO2 by carbon isotope composition[J]. Chinese Journal of Environmental Science, 2012, 33(4): 1041-1049.(in Chinese). doi:
  3. 3.
    YUAN F, GAO J, YAO L, et al. Development of highly sensitive balloon-borne methane measurement system based on cavity ringdown spectroscopy[J]. Opt. Precision Eng., 2020, 28(9): 1881-1892.(in Chinese). doi:
  4. 4.
    AI Y K, LI J, LI Q Y, et al. Cavity ringdown spectroscopy of nitric oxide in the ultraviolet region for human breath test[J]. Journal of Breath Research, 2020, 14(3): 037101. doi:
  5. 5.
    MONDAL K, KAIPARA R, RAJAKUMAR B. Investigation of the absorption cross section of phenyl radical and its kinetics with methanol in the gas phase using cavity ring-down spectroscopy and theoretical methodologies[J]. The Journal of Physical Chemistry A, 2019, 123(45): 9682-9692. doi:
  6. 6.
    WANG J N, ZHENG L J, NIU X T, et al. Design and fabrication of a mid-infrared carbon dioxide sensor for the application in greenhouse environment[J]. Optoelectronics Letters, 2016, 12(5): 379-383. doi:
  7. 7.
    ESLER M B, GRIFFITH D W T, WILSON S R, et al. Precision trace gas analysis by FT-IR spectroscopy. 1. simultaneous analysis of CO2, CH4, N2O, and CO in air[J]. Analytical Chemistry, 2000, 72(1): 206-215. doi:
  8. 8.
    LI M X, CHEN B, RUAN J, et al. On-line detection of carbon dioxide in large scale offshore by laser technology[J]. Opt. Precision Eng., 2020, 28(7): 1424-1432.(in Chinese). doi:
  9. 9.
    FUNKE H H, GRISSOM B L, MCGREW C E, et al. Techniques for the measurement of trace moisture in high-purity electronic specialty gases[J]. Review of Scientific Instruments, 2003, 74(9): 3909-3933. doi:
  10. 10.
    ROMANINI D, KACHANOV A A, SADEGHI N, et al. CW cavity ring down spectroscopy[J]. Chemical Physics Letters, 1997, 264(3/4): 316-322. doi:
  11. 11.
    PALDUS B A, JrHARRIS J S, MARTIN J, et al. Laser diode cavity ring-down spectroscopy using acousto-optic modulator stabilization[J]. Journal of Applied Physics, 1997, 82(7): 3199-3204. doi:
  12. 12.
    HAHN J W, YOO Y S, LEE J Y, et al. Cavity ringdown spectroscopy with a continuous-wave laser: calculation of coupling efficiency and a new spectrometer design[J]. Applied Optics, 1999, 38(9): 1859-1866. doi:
  13. 13.
    ORR BJ, HE Y. Rapidly swept continuous-wave cavity-ringdown spectroscopy[J]. Chemical Physics Letters, 2011, 512(1/2/3): 1-20. doi:
  14. 14.
    CHEN B, SUN Y R, ZHOU Z Y, et al. Ultrasensitive, self-calibrated cavity ring-down spectrometer for quantitative trace gas analysis[J]. Applied Optics, 2014, 53(32): 7716. doi:
  15. 15.
    CROSSON E R. A cavity ring-down analyzer for measuring atmospheric levels of methane, carbon dioxide, and water vapor[J]. Applied Physics B, 2008, 92(3): 403-408. doi:
  16. 16.
    WANG J D, YU J, MO Z Q, et al. Multicomponent gas detection based on concise CW-cavity ring-down spectroscopy with a bow-Tie design[J]. Applied Optics, 2019, 58(11): 2773. doi:
  17. 17.
    SONG S M. Continuous Wave Cavity Ring-Down Spectroscopy Device and Trace Methane Detection[D]. Changchun: Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, 2019. (in Chinese)
  18. 18.
    GUO R M, TENG J H, CAO K, et al. Comb-assisted, Pound-Drever-Hall locked cavity ring-down spectrometer for high-performance retrieval of transition parameters[J]. Optics Express, 2019, 27(22): 31850. doi:
  19. 19.
    MA G S, LIU Y, DENG H, et al. High finesse optical feedback cavity ringdown spectroscopy[J]. Opt. Precision Eng., 2022, 30(19): 2305-2312. (in Chinese). doi:
  20. 20.
    BURKART J, ROMANINI D, KASSI S. Optical feedback frequency stabilized cavity ring-down spectroscopy[J]. Optics Letters, 2014, 39(16): 4695. doi:
  21. 21.
    HU M, CHEN X, ZHANG H, et al. Fast optical cavity ring-down spectroscopy detection based on first harmonic frequency locking[J]. Opt. Precision Eng., 2022, 30(4): 363-371. (in Chinese). doi:

Читать полностью