MI Bao-Yong . Photoelectric Automatic Refractometer with HighPrecision in the Wavelength Range 365 to 12000nm[J]. Editorial Office of Optics and Precision Engineering, 1998,(3): 79-84
MI Bao-Yong . Photoelectric Automatic Refractometer with HighPrecision in the Wavelength Range 365 to 12000nm[J]. Editorial Office of Optics and Precision Engineering, 1998,(3): 79-84DOI:
Photoelectric Automatic Refractometer with HighPrecision in the Wavelength Range 365 to 12000nm
This paper describes measuring principle of the photo electric automatic refractometer with high precision in the wavelength range 365 to 12000nm and its constraction and discusses the sources of error.The refractive indices of different optical materials are measured with this instrument.The measuring uncertainies of better than 3?10
-6
and 3?10
-5
in the wavelength range of 365 to 2600 nm and 2600 to 12000 nm are obtained
Sensing properties of ellipsoidal extrinsic Fabry-Perot fiber probe
Laser confocal interference surface rapid and high-precision automatic measurement
Automatic measurement of center deviation of lens of laser differential confocal
Particle sorting with different refractive indices using single fiber optical tweezers
Measurement of all-fiber multi-feedback liquid refractive index
Related Author
ZHAO Mingfu
CHEN Jian
ZOU Xue
LUO Binbin
SHI Shenghui
LI Gongye
ZHAO Jiahao
YANG Shuai
Related Institution
Chongqing Key Laboratory of Optical Fiber Sensor and Photoelectric Detection, Chongqing University of Technology
MIIT Key Laboratory of Complex-field Intelligent Exploration,School of Optics and Photonics,Beijing Institute of Technology
Department of Electrical Information Engineering, Northeast Petroleum University, Qinhuangdao Campus
College of Electrical Information Engineering, Northeastern Petroleum University
Key Laboratory of the Ministry of Education for Optoelectronic Measurement Technology and Instrument, Beijing Information Science & Technology University