Correcting method for ring artifacts in fan-beam X-ray ICT
the non-uniform response of the detecting system could cause ring artifacts in image reconstruction. In order to mitigate this situation
a class of scan sampling patterns was proposed
which could suppress the non-uniform response and consequently decrease the ring artifacts. A condition
which this class of scan patterns must meet
is discussed. Computer simulation and experimental results show that the reconstruction of image with few ring artifacts
and without any other special disposal
could be generated with a filtered back-projection algorithm using those data sampled by the detecting system with non-uniform response under this kind of scan pattern.
关键词
Keywords
references
Horn B K P. Fan-Beam Reconstruction Methods[J]. IEEE, 1979, 67: 1616-1623.
Herman G T. Image reconstruction from projection: the fundamentals of computed tomography[M]. New York: Academic Press, 1980.
庄天戈. CT原理与算法[M]. 上海: 上海交通大学出版社, 1992. 68-71.
Kowalski G. Suppression of ring artifacts in CT fan-beam scanners[J]. IEEE Trans On Nucl Sci, 1978,NS 25(5): 1111-1116.
Jiang Hesieh. Reconstruction Bias Resulting from Weighted Projection and Iso-center Misalignment[J]. SPIE, 1999, 3661: 442-449.
Davis G R, Elliott J C. X-ray Microtomography Scanner Using Time-delay Integration For Elimination of Ring Artifacts in the Reconstructed Image[J]. Nucl. Instr. And Meth. in Phys.Res. A, 1998, 402:683-687.