

浏览全部资源
扫码关注微信
1. Optical and Electronic Information Engineering College,University of Shanghai for Science &amp
2. Institute for Laser Medicine &amp
3. Technology Shanghai,China,200093
4. Bio-photonics,Shanghai Jiaotong University Shanghai,China,200030
Received:16 March 2005,
Revised:27 June 2005,
Published Online:30 October 2005,
Published:30 October 2005
移动端阅览
LIU Yong, CHEN Jia-bi, REN Qiu-shi, et al. Axial intensity in a fiber-optic confocal microscope[J]. Optics and precision engineering, 2005, 13(5): 608-612.
A fiber-optic confocal microscope has been analyzed by Fourier optics.It is found that the detected light intensity has three parts
each of which is depennted on the coupled lens
the detective lens
and the part comprised of the fiber and the microprobe.The simulated results show that the less the value of the parameter
A
is
which is dependent on the fiber and microprobe
the higher the axial resolution of the system is. For the case
as
A
→∞
the axial resolution is zero
which is corresponding to the conventional microscope.as
A
≤1
the axial resolution changes slightly
and is close to the optimal value
which is corresponding to the perfect confocal microscope.when the reflective loss takes place at the end of fiber
the contrast of axial intensity will decrease.All that will help the design of endoscope with confocal microscope at cellular level.
A fiber-optic confocal microscope has been analyzed by Fourier optics.It is found that the detected light intensity has three parts
each of which is depennted on the coupled lens
the detective lens
and the part comprised of the fiber and the microprobe.The simulated results show that the less the value of the parameter
A
is
which is dependent on the fiber and microprobe
the higher the axial resolution of the system is. For the case
as
A
→∞
the axial resolution is zero
which is corresponding to the conventional microscope.as
A
≤1
the axial resolution changes slightly
and is close to the optimal value
which is corresponding to the perfect confocal microscope.when the reflective loss takes place at the end of fiber
the contrast of axial intensity will decrease.All that will help the design of endoscope with confocal microscope at cellular level.
. SABHARWAL Y S,ROUSE A R.Slit-scanning confocal microendoscope for high-resolution in vivo imaging[J].Appl.Opt.,1999,38:7133-7144.
. KNITTEL J.Endoscope-compatible confocal microscope using a gradient index-lens system[J].Opt.Commu,2001,188:267-273
. GMITRO A F.Confocal microscopy through a fiber-optic imaging bundle[J].Opt.lett.,1993,18:565-567
. SUNG K B.Near real time in vivo fiber optic confocal microscopy:sub-cellular structure resolved[J]. J.Micros,2002,207:137-145.
. WU CH Q.Optical waveguide theory[M].Tsinghua University press,:80-82.
. GU M,GAN X,SHEPPARD C J P.Three-dimensional coherent transfer functions in fiber-optical confocal scanning microscopes[J].J.Opt.Soc.A.,1991,7:1019-1025.
0
Views
271
下载量
1
CSCD
Publicity Resources
Related Articles
Related Author
Related Institution
京公网安备11010802024621