A new mini-fiber-spectrometer adopting micro-silicon-slit
A new mini-fiber-spectrometer was developed with micro-silicon-slit instead of traditional mechanical slit. By calculating the impact of nonlinearity MEMS-based slit on the resolution of the mini-fiber-spectrometer and measuring the resolution of the system
it finds that using micro-silicon slit is efficient. Meantime
a theory for wavelength-pixel calibrating was brought forward after analyzing the resolution of pixel and FWHM
which was validated by analyzing data of the experiments of measuring the mercury-arc and fine calculating through the tool of MATLAB system. The results show that the new mini-fiber-spectrometer with micro-silicon-slit is worked out
its FWHM is 0.85 nm
the precision of the wavelength calibration is less than 0.2 nm and the volume is 50 mm×46 mm×14 mm.
关键词
Keywords
references
. 于常青,刘杰,李家泽,等. 光学多通道分析仪的分辨率分析[J]. 光学技术,2001,27(2):97-99. YU CH Q, LIU J, LI J Z,et al. The resolution of optic multichannel analyzer[J].Optical Technique,2001,27(2):97-99.(in Chinese)
. 鞠挥,吴一辉. 微型光谱仪的发展现状[J]. 光学 精密工程,2001,9(4):372-376. JU H,WU Y H.The State of the art of micro spectrometers[J].Optics and Precision Engineering,2001,9(4):372-376.(in Chinese)
. 赵育良, 张青臣, 李开端. 基于线阵CCD 的全息光栅位移传感器[J]. 测试技术学报, 2002,16(3):167-170. ZHAO Y L,ZHANG Q CH,LI K D.The holographic grating displacement sensors based on linear CCD[J].Journal of Test and Measurement Technology,2002,16(3):167-170.(in Chinese)
. 李昌厚. 光谱带宽与分析测试误差的关系[J]. 分析测试技术与仪器,2004,10(2):65-67. LI CH H. Study on relationship between the spectrum band width and the absorbance error[J].Analysis and Testing Technology and Instruments,2004,10(2):65-67.(in Chinese)
. 鞠挥.用于生化分析的光谱仪微小型化的研究 . 中国科学院长春光学精密机械与物理研究所,2002. JU H.Study on miniaturization of spectrometer for biochemical use . Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences,2002.(in Chinese)
. 胡松,温志渝,等. 一种新型微小型光谱仪的设计 . 压电与声光,2000,22(6):363-366. HU S, WEN ZH Y,et al. Design of a novel mini-spectrometer[J]. Piezoelectrics & Acoustooptics,2000,22(6):363-366.(in Chinese)
. 徐中民,禹秉熙. 对PC2000-PC/104型光谱仪的波长定标[J]. 光学 精密工程,2004,12(1):11-14. XU ZH M, YU B X. Wavelength calibration for PC2000-PC/104 spectrometer[J].Optics and Precision Engineering,2004,12(1):11-14.(in Chinese)
. 费业泰.误差理论与数据处理[M]. 北京:机械工业出版社,2000. FEI Y T. Eerror theory and data processing[M].Beijing: Mechanical Industry Press,2000.(in Chinese)