Improvement of spectral domain OCT image quality by linear interpolation and CCD response compensation
|更新时间:2020-08-12
|
Improvement of spectral domain OCT image quality by linear interpolation and CCD response compensation
Optics and Precision EngineeringVol. 14, Issue 5, Pages: 929-933(2006)
作者机构:
1. 天津大学 精密仪器与光电子工程学院 天津,300072
2. 金日成综合大学 物理学院,朝鲜
作者简介:
基金信息:
DOI:
CLC:O433.4;O436.1
Received:14 February 2006,
Revised:18 August 2006,
Published Online:30 October 2006,
Published:30 October 2006
稿件说明:
移动端阅览
ZHANG Tae-Sok, LI Gang, ZHENG Yu, et al. Improvement of spectral domain OCT image quality by linear interpolation and CCD response compensation[J]. Optics and precision engineering, 2006, 14(5): 929-933.
DOI:
ZHANG Tae-Sok, LI Gang, ZHENG Yu, et al. Improvement of spectral domain OCT image quality by linear interpolation and CCD response compensation[J]. Optics and precision engineering, 2006, 14(5): 929-933.DOI:
Improvement of spectral domain OCT image quality by linear interpolation and CCD response compensation
Data collected by CCD camera is a function of wavelength in Spectral-Domain Optical Coherence Tomography (SDOCT) system. By consideration of a wavelength-depended response characteristics of CCD and the Fourier transform relationship between time (distance) and frequency (wave-number)
the CCD data must be responsivity-compensated and resampled by linear interpolation in frequency space. Based on the above analysis
the CCD response compensation method and the linear interpolation arithmetic were suggested and tested by the sample imaging. Results show that signal-to-noise ratio (SNR) increases a 30% in proposed method and arithmetic mentioned above