LI Hongyu,GUO Hanzhou,LI Junlin,et al.Transmittance measurement device of large-aperture infrared optical system[J].Optics and Precision Engineering,2023,31(12):1752-1760.
To solve the transmittance measurement problem of large-aperture infrared optical systems in complete machines, a single-channel infrared transmittance measurement method based on grayscale images is proposed, and a transmittance measurement device is established. First, a blackbody and a target are placed at the focal plane of the infrared optical system under measurement according to the specified position relationship, and an infrared thermal imager is placed at the entrance port of the infrared optical system to capture the radiation image. Then, the infrared thermal imager is placed in front of the blackbody to capture the radiation image of the blackbody. Finally, the mean value of the radiation image and the mean value of the blackbody radiation image are calculated; the ratio of the two values is the transmittance of the measured system. Upon measuring the infrared transmittance of a large-aperture infrared optical system by using the proposed method, the results indicate that the method can accurately measure the transmittance of large-aperture infrared optical systems. The absolute error between the measured value and the design value is 0.85%. Analyzing the uncertainty of the test results reveals that the transmittance measurement accuracy of the proposed method is approximately 1.04%. This method entails a simple measurement process and is effective for evaluating the transmittance of large-aperture infrared optical systems in complete machines.
LIU H X , LI X S , SUN J X , et al . Haze and transmittance real-time tester for curved-surface and transparent samples [J]. Opt. Precision Eng. , 2018 , 26 ( 12 ): 2881 - 2887 . (in Chinese) . doi: 10.3788/ope.20182612.2881 http://dx.doi.org/10.3788/ope.20182612.2881
TIAN L D , WANG T , ZHAO H X , et al . Infrared optical system spectrum transmittance testing device and method : CN109060731B [P]. 2020-04-03 . (in Chinese) .
LIU J C , LI H W , WANG J L , et al . Fast radiance calibration for ground-based large-aperture infrared opto-electric equipment [J]. Acta Optica Sinica , 2015 , 35 ( 3 ): 17 - 26 . (in Chinese) . doi: 10.3788/aos201535.0301003 http://dx.doi.org/10.3788/aos201535.0301003
HUANG Z G . Research on Infrared Detection Technology of Space Target Ground [D]. Changchun : Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences , 2018 . (in Chinese)
ZHANG X J , FAN Y C , BAO H , et al . Applications and development of ultra large aperture space optical remote sensors [J]. Opt. Precision Eng. , 2016 , 24 ( 11 ): 2613 - 2626 . (in Chinese) . doi: 10.3788/OPE.20162411.2613 http://dx.doi.org/10.3788/OPE.20162411.2613
JI Y L , HAO X P , SUN Y D , et al . Research on large-area blackbody radiation source for infrared remote sensor calibration [J]. International Journal of Thermophysics , 2022 , 43 ( 9 ): 1 - 16 . doi: 10.1007/s10765-022-03067-0 http://dx.doi.org/10.1007/s10765-022-03067-0
王立秋 . 光谱透过率测试系统设计及优化 [D]. 长春 : 长春理工大学 , 2016 .
WANG L Q . Design and Optimization of Spectral Transmittance Testing System [D]. Changchun : Changchun University of Science and Technology , 2016 . (in Chinese)
LI Y , TIAN M , WANG J S , et al . Infrared transmittance and error factor analysis of germanium glass measured by thermal imager [J]. Journal of Changchun University of Science and Technology , 2022 , 45 ( 6 ): 79 - 83 . (in Chinese)
ROGALSKI A . Infrared Detectors , Second Edition [M]. Florida : CRC Press , 2020 . doi: 10.1201/b21951 http://dx.doi.org/10.1201/b21951
YANG G Q . Research on Key Technologies for Improving Measurement Accuracy of Ground-Based Infrared Radiation Measurement System [D]. Changchun : Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences , 2020 . (in Chinese)
GUTSCHWAGER B , TAUBERT D , HOLLANDT J . Analysis of reference sources for the characterization and calibration of infrared cameras [J]. International Journal of Thermophysics , 2015 , 36 ( 2 ): 303 - 314 . doi: 10.1007/s10765-014-1767-1 http://dx.doi.org/10.1007/s10765-014-1767-1
LIU T J , ZHANG T , LI J L , et al . Effect of spatial distribution of test device on noise equivalent temperature difference test of infrared camera [J]. Laser & Optoelectronics Progress , 2023 , 60 ( 6 ): 170 - 174 . (in Chinese)
费业泰 . 误差理论与数据处理 [M]. 7版 . 北京 : 机械工业出版社 , 2015 .
FEI Y T . Error Theory and Data Processing [M]. 7th ed . Beijing : China Machine Press , 2015 . (in Chinese)