PAN Ming-hua, WEN Xiang-wen, ZHU Guo-li. Design of combination measurement system for pitching angles[J]. Editorial Office of Optics and Precision Engineering, 2011,19(3): 598-604
PAN Ming-hua, WEN Xiang-wen, ZHU Guo-li. Design of combination measurement system for pitching angles[J]. Editorial Office of Optics and Precision Engineering, 2011,19(3): 598-604 DOI: 10.3788/OPE.20111903.0598.
Design of combination measurement system for pitching angles
A combination measurement system for pitching angles was put forward based on a collimated laser
a single-axis inclinator and a CCD camera to improve the precision and stability of measuring pitching angles in engineering. The combination measurement model was introduced and its combination measurement principle was studied in detail on the basis of the coordinate rotation algorithm and the iteration method. The global sensitive analysis of each factor affecting the combination measurement was carried out by using Monte Carlo method.Obtained results show that compared with the direct measurement method with an inclinator widely used in pitching angle measuring at present
the proposed method can obtain much higher measurement precision and stability. A measurement experiment on a highly accurate three-axis turntable was performed
which indicates that the maximum absolute error is 0.3 mrad and the standard deviation is 0.16 mrad.These data are better than that from the direct measurement method with the inclinator
and meet the requirements of the engineering applications.
关键词
Keywords
references
KIRKPATRICK S J, DUNCAN D D. Noncontact microstrain measurements in orthodontic wires [J]. Journal of Biomedical Materials Research, 1995,29(11):1437-1442.[2] LIU D, WANG F, CEN K F, et al.. Noncontact temperature measurement by means of CCD cameras in a participating medium [J]. Optics Letters, 2008, 33(5):422-424.[3] TANG W SH, LI Q H, TAN Q CH, et al.. A new method of noncontact measurement of the plane parameters . 2009 IEEE International Conference on Mechatronics and Automation. 445 Hoes Lane-P.O.Box 1331, Piscataway, NJ 08855-1331, United States: IEEE Computer Society, 2009:2554-2559.[4] 郝鹏飞, 张新军, 董志国, 等. 基于面阵CCD的高精度测量技术及应用 [J]. 天津大学学报, 2008, 41(12):1443-1447. HAO P F, ZHANG X J, DONG ZH G, et al.. Area CCDbased high accuracy measurement technology and its application [J]. Journal of Tianjin University, 2008,41(12):1443-1447. (in Chinese)[5] 李 珺 , 姚进, 任违. 天平校准系统中位置姿态非接触式测量与计算方法 [J]. 四川大学学报(工程科学版), 2003,35(3):90-92. LI J, YAO J, REN W. Non-contacted measure to confirm the balances position in adjustment of balance [J]. Journal of Sichuan University (Engineering Science Edition), 2003,35(3):90-92. (in Chinese)[6] 李慧鹏, 王军, 张军, 等. 一种新型非接触式线径测量系统 [J]. 仪器仪表学报, 2004,25(4):209-211. LI H P, WANG J, ZHANG J, et al.. A novel noncontact measurment system for the diameter of wires [J]. Chinese Journal of Scientific Instrument, 2004,25(4):209-211. (in Chinese)[7] 艾莉莉, 袁峰, 丁振良. 应用线阵CCD 的空间目标外姿态测量系统 [J]. 光学 精密工程, 2008,16(1):161-165. AI L L, YUAN F, DING ZH L. An exterior attitude measurement system for spatial object based on linear CCD [J]. Opt. Precision Eng., 2008, 16 (1):161-165. (in Chinese)[8] SRIKANTA M,NEIL D,RUSKAUFF G. Global sensitivity analysis techniques for probabilistic ground water modeling [J]. Ground Water, 2009,47(5):730-747.[9] STEFANO C,LOVATO T,RATTO M. Global uncertainty and sensitivity analysis of a food-web bioaccumulation model [J]. Environmental Toxicology and Chemistry, 2009,28(4):718-732.[10] 邓兆祥,颜长征. 基于蒙特卡罗法的摩托车车架灵敏度分析[J]. 重庆大学学报,2008,31(10):1113-1117. DENG ZH X, YAN CH ZH. Sensitivity analysis of a motorcycle frame using the Monte Carlo method [J]. Journal of Chongqing University, 2008,31(10):1113-1117.(in CHinese)[11] 于德介,李睿. Sobol法在非线性被动隔振体灵敏度分析中的应用研究[J]. 振动工程学报,2004,17(2):200-213. YU D J, LI R. Application of Sobol method to sensitivity analysis of a non-linear passive vibration isolators [J]. Journal of Vibration Engineering, 2004,17(2):200-213.(in Chinese)[12] 陈刚,汪玉,毛为民,等. 冲击载荷作用下舰艇管路系统全局参数灵敏度分析[J]. 振动与冲击,2006,26(3):45-48. CHEN G, WANG Y, MAO W M, et al.. Global parameter sensitivity analysis of shipboard piping systems under shock loads [J]. Journal of Vibration and Shock, 2006,26(3):45-48. (in Chinese)