Interstitial fluid transdermal extraction tool based on microfluidics technology
A microfluidic based Interstitial Fluid (ISF) transdermal extraction tool made from polydimethylsiloxane (PDMS) was designed to the application of continuous glucose monitoring. Four layers of PDMS fabricated by replica molding process were bonded together using oxygen plasma to construct the device
including a Venturi tube for vacuum generation
chambers for the introduction of ISF and normal saline solution
pneumatic valves for fluid control
and interconnected microchannels. Then
the output vacuum of Venturi tube and the close pressure of pneumatic valve were measured. Finally
the ISF extraction and collection function was tested. Results show that the vacuum pressure is used for fluid manipulation and a 92 kPa (absolute pressure) vacuum is achieved
when 220 kPa (absolute pressure) external pressure is applied to the Venturi tube. Moreover
the normally open pneumatic valves can be closed under the operating pressure less than 65 kPa (relative pressure). Under the control of pneumatic valves
the normal saline injection
ISF extraction and collection function of the tool are implemented by using the vacuum generated from the Venturi tube.
关键词
Keywords
references
KLONOFF D C. A review of continuous glucose monitoring technology [J]. Diabetes Technology & Therapeutics, 2005,7(5):770-775.[2] MITRAGOTRI S, COLEMAN M, KOST J, et al.. Analysis of ultrasonically extracted interstitial fluid as a predictor of blood glucose levels [J]. Journal of Applied Physiology, 2000,89(3):961-966.[3] TIESSEN R G, RHEMREV-BOOM M M, KORF J. Glucose gradient differences in subcutaneous tissue of healthy volunteers assessed with ultraslow microdialysis and a nanolitre glucose sensor [J]. Life Sciences, 2002, 70(21):2457-2466.[4] GARG S K, POTTS R O, ACKERMAN N R, et al.. Correlation of fingerstick blood glucose measurements with GlucoWatch biographer glucose results in young subjects with type 1 diabetes [J]. Diabetes Care, 1999,22(10):1708-1714.[5] GUERCI B, FLORIOT M, BOHME P, et al.. Clinical performance of CGMS in type 1 diabetic patients treated by continuous subcutaneous insulin infusion using insulin analogs [J]. Diabetes Care, 2003,26(3):582-589.[6] BURGE M R, MITCHELL S, SAWYER A, et al.. Continuous glucose monitoring: the future of diabetes management [J]. Diabetes Spectrum, 2008,21(2):112-119.[7] BECK R W, HIRSCH I B, LAFFEL L, et al.. The effect of continuous glucose monitoring in well-controlled type 1 diabetes [J]. Diabetes Care, 2009,32(8):1378-1383.[8] TAMBORLANE W V, BECK R W, BODE B W, et al.. Continuous glucose monitoring and intensive treatment of type 1 diabetes [J]. The New England Journal of Medicine, 2008,359(14):1464-1476.[9] MITRAGOTRI S. Synergistic effect of enhancers for transdermal drug delivery [J]. Pharmaceutical Research, 2000,17(11):1354-1359.[10] KOST J, MITRAGOTRI S, GABBAY R A, et al.. Transdermal monitoring of glucose and other analytes using ultrasound [J]. Nature Medicine, 2000,6:347-350.
Minimally invasive continuous blood glucose monitor based on microfluidic and enzyme colorimetric technologies
Minimally-invasive blood glucose detection instrument based on surface plasmon resonance sensor decorated with GGBP
Related Author
Ke-xin XU
Xiao-long CHEN
Da-chao LI
Hai-xia YU
LI Da-chao
LI Guo-qing
ZHANG Jing-xin
YU Hai-xia
Related Institution
State Key Laboratory of Precision Measuring Technology and Instruments, School of Precision Instrument and Optoelectronics Engineering, Tianjin University