REN Jian-wei, ZHANG Yan-qi, YE Zhao, QUAN Xian-rong. Establishment and application of CCD radiation response function matrix[J]. Editorial Office of Optics and Precision Engineering, 2012,20(5): 957-962
REN Jian-wei, ZHANG Yan-qi, YE Zhao, QUAN Xian-rong. Establishment and application of CCD radiation response function matrix[J]. Editorial Office of Optics and Precision Engineering, 2012,20(5): 957-962 DOI: 10.3788/OPE.20122005.0957.
Establishment and application of CCD radiation response function matrix
To describe the parameters of a big format Charge Coupled Device (CCD) by a function with definit physical meaning and to evaluate its radiation response performance directly
a concept of "radiation response function matrix" was proposed to describe the radiation response parameters of each pixel of the CCD. Firstly
by analyzing the physical concept of each element
some parameters such as absolute radiate response
nonlinearity
dark noise
Signal-to-Noise Ratio (SNR) and the non-uniformity of each pixel of the CCD were characterized by the matrix. Then
response characteristic coefficient of each pixel was obtained using the model of multiple regression analysis by a test on the area CCD KAI-16000. Finally
by taking the result for an example
application of this matrix was discussed. Experimental results show that the response of this CCD is linear
and its non-uniformity is 3.1%
dark noise is 3.84. These results demonstrate that the method is feasible and practical
and it can satisfy the requirement of the description of a big format CCD.
关键词
Keywords
references
韩心志,焦世举.航天光学遥感辐射度学[M]. 哈尔滨:哈尔滨工业大学出版社 ,1994. HAN X ZH, JIAO SH J. Space Optics Remote Sensing Radiance[M]. Harbin: Harbin Institute of Technology Press, 1994. (in Chinese)[2] 胡君,王栋. 空间光学遥感器的多光谱TDICCD信号检测与生成 [J]. 光学 精密工程,2009,8(11): 1810-1818. HU J, WANG D. Detection and generation of multi-spectral TDICCD signals for space optical remote sensors[J]. Opt. Precision Eng., 2009, 8(11): 1810-1818.(in Chinese)[3] 叶培建,刘福安,曹海翊. 线阵CCD器件主要性能参数及测试方法 [J]. 中国空间科学技术,1997,6(3): 44-51. YE P J, LIU F A, CAO H Y. The main performances and test method of the linear charge coupled device[J]. Chinese Space Science and Technology, 1997, 6(3): 44-51. (in Chinese)[4] 奕中杰. 线阵CCD光电参量测试系统的研究 [J]. 光学 精密工程,1994,2(4): 140-149. LUAN ZH J. Study a test system of linear CCD photoelectronic parameters[J]. Opt. Precision Eng., 1994, 2(4): 140-149.(in Chinese)[5] 任建伟,万志,李宪圣,等. 空间光学遥感器的辐射传递特性与校正方法 [J]. 光学 精密工程,2007,15(8):1186-1190. REN J W, WAN ZH, LI X SH, et al.. Radiation transfer characteristic and calibrating method for space optical remote sensor[J]. Opt. Precision Eng., 2007, 15(8): 1186-1190. (in Chinese)[6] 陈风,郑小兵. 光谱非匹配对光学遥感器定标精度的影响 [J].光学 精密工程,2008, 16(3): 415-419. CHEN F,ZHENG X B. Influence of spectrum not-matching on calibration precision of remote sensor[J].Opt. Precision Eng., 2008,16(3): 415-419. (in Chinese)[7] 任建伟,刘则洵,万志,等. 离轴三反宽视场空间相机的辐射定标 [J].光学 精密工程,2010, 18(7): 1491-1497. REN J W, LIU Z X,WAN ZH, et al.. Radiometric calibration of off-axis three-mirror-anastigmat space camera with wide viewing field[J].Opt. Precision Eng., 2010,18(7): 1491-1497.(in Chinese)[8] 崔敦杰. 光学遥感器的辐射定标纪念建所五十周年论文及论文摘要汇编(一) .长春:长春光学精密机械与物理研究所,2002: 148-154. CUI D J. Radiometric calibration of optical remote sensor .Changchun: Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences.2002: 148-154.(in Chinese)[9] 张立国,李豫东,刘则洵,等. TDI-CCD总剂量辐射效应及测试 [J]. 光学 精密工程,2009,17(12):2924-2930. ZHANG L G, LI Y D, LIU Z X, et al.. Influence of total dose effects on TDI-CCD and corresponding test methods[J]. Opt. Precision Eng., 2009, 17(12): 2924-2930. (in Chinese)[10] 车念曾,闫达远. 辐射度学和光度学[M]. 北京:北京理工大学出版社 ,1990. CHE N Z, YAN D Y. Radiance and Luminosity[M]. Beijing: Beijing Institute of Technology Press, 1990.(in Chinese)