SONG Mao-xin, SUN Bin, SUN Xiao-bing, HONG Jin. Polarization calibration of airborne muti-angle polarimetric radiometer[J]. Editorial Office of Optics and Precision Engineering, 2012,20(6): 1153-1158
SONG Mao-xin, SUN Bin, SUN Xiao-bing, HONG Jin. Polarization calibration of airborne muti-angle polarimetric radiometer[J]. Editorial Office of Optics and Precision Engineering, 2012,20(6): 1153-1158 DOI: 10.3788/OPE.20122006.1153.
Polarization calibration of airborne muti-angle polarimetric radiometer
The polarization calibration of an airborne multi-angle polarimetric radiometer was researched to eliminate the polarization effects induced by itself and to improve its measurement accuracy. First
instrument detection equations with calibration coefficients were derived according to the main factors that have been induced into the instrumental polarization effects
and the expression of calibration coefficients was solved based on the detection equations for the same signal when the instrument was fixed at states of 0癮nd 90?.Then
the calibration method to get the signals of the two instrument states and to solve the calibration coefficients were designed. Finally
the polarization calibration coefficients were computed from the measured data using a non-polarized light source and a completely linearly polarized one
respectively
and the polarized calibration results were verified by adjusting the polarization degree of the incident light. The results indicate that the instrument can offer a polarized measurement accuracy better than 0.5%
which meets the requirements of the instrument polarimetry.
关键词
Keywords
references
TRAVIS L D. Earth Oberving Scanning Polarimeter EOS Reference Handbook[M]. NASA, Washington DC, 1995:77-129.
CAIRNS B,EDGAR E,RUSSEL L,et al..The research scanning polarimeter: calibration and ground-based measurements[J].SPIE,1999,3754:186-196.
RICHARD J, NARDELL P C,CAIRNS B, et al..Aerosol Polarimetry Sensor for the Glory Mission [J].SPIE,2007,6786:67865L-1-L-17.
GODDARD SPACE FLIGHT CENTER.Aerosol polarimetry sensor calibration .Glory Project,2010.
THIERRY B D,YVES A,LAHERRERE J M,et al.. Pre-flight calibration of the POLDER instrument[J].SPIE,1993,2553:218-231.
顾行发,陈兴峰,程天海. 多角度偏振遥感相机DPC在轨偏振定标[J]. 物理学报,2011,60(7) :070702-1-070702-8. GU X F,CHEN X F,CHENG T H.In-flight polarization calibration methods of directional polarized remote sensing camera DPC[J]. Acta Phys.Sin., 2011,60(7): 070702-1-070702-8. (in Chinese)
陈立刚,孟凡刚,袁银麟,等. 偏振相机的光学定标方案研究[J]. 大气与环境光学学报,2010,5(3):126-231. CHEN L G,MENG F G,YUAN Y L, et al..Calibration method for polarization camera[J].Journal of Atmospheric and Envirmental Optics,2010,5(3):126-231.(in Chinese)
陈立刚,洪津,乔延利,等. 非理想正交反射镜消偏性能的模拟研究[J]. 应用光学,2008,29(4):633-638. CHEN L G, HONG J , QIAO Y L, et al.. Simulation study on depolarization for imperfect orthogonal mirrors[J].Journal of Applied Optics,2008,29(4):633-638. (in Chinese)
李幼平,禹秉熙,王玉鹏. 成像光谱仪辐射定标影响量的测量链与不确定度[J]. 光学 精密工程,2006,14(5):822-827. LI Y P, YU B X,WANG Y P.Measurement chain of influence quantities and uncertainty of radiometric calibration for imaging spectrometer[J].Opt. Precision Eng.,2006,14(5):822-827.(in Chinese)
陈立刚,洪津,乔延利. 一种高精度偏振遥感探测方式的精度分析[J]. 光谱学与光谱分析,2008,28(10):2384-2387. CHEN L G, HONG J, QIAO Y L.Accuracy analysis on a sort of polarized measurement in remote sensing[J].Spectroscopy and Spectral Analysis, 2008,28(10):2384-2387. (in Chinese)
LABORATOIRE D’OPTIQUE ATMOSPHERIQUE-Universite des Sciences et Technologies de Lille. Polarizing Box Polbox User’s Guide .1998.
廖延彪.偏振光学[M]. 北京:科学出版社,2003,45-63. LIAO T B.Polarization Optics [M].Beijing :Science Press,2003.45-63.(in Chinese)