ZHAO En-ming, LUO Shen-zi, YUAN Li-bo, YANG Yuan-yuan, PENG Li-rong, YANG Xing-hua. Preparation and performance testing of fiber-optic O<sub>2</sub> sensitive probe[J]. Editorial Office of Optics and Precision Engineering, 2012,20(11): 2411-2415
ZHAO En-ming, LUO Shen-zi, YUAN Li-bo, YANG Yuan-yuan, PENG Li-rong, YANG Xing-hua. Preparation and performance testing of fiber-optic O<sub>2</sub> sensitive probe[J]. Editorial Office of Optics and Precision Engineering, 2012,20(11): 2411-2415 DOI: 10.3788/OPE.20122011.2411.
Preparation and performance testing of fiber-optic O2 sensitive probe
An oxygen optical sensing probe was fabricated by modifying the sensitive film doped with Ru compounds in a holy optical fiber. The structure of this probe was based on the simple holy optical fiber with internal microchannel structure throughout the whole fiber
in which the microchannel structure could be used as a carrier of sensitive materials and a micro-sensing pool and the sensing process was finished in it. The sensing film was deposited in the optical fiber by inhaling the cellulose solution containing [Ru(dpp)
3
]
2+
directly. Furthermore
the sensing process was based on fluorescence quenching process. As the oxygen could strongly quench the fluorescence of [Ru(dpp)
3
]
2+
in the sensitive film
the sensing probe could exhibit different fluorescence intensities for different oxygen concentrations. The experimental results show that the designed probe responses to the oxygen in linearity for a full concentration range of 0~100%
and the ratio of I
0
/I
100
(where I
0
and I
100
respectively represent the fluorescence intensities of the optrode exposed to 100% N
2
and 100% O
2
) is 9.4. Simultaneously
the probe can make a quick response within 50 ms. The experiment also show that the error between experimental value and real value is less than 3% and the resolution can be 2% in a normal pressure.
关键词
Keywords
references
魏茂金,赖森财,黄思俞,等. 磁动式氧含量分析仪的研制[J]. 光学 精密工程, 2007, 15(5):761-766. WEI M J, LAI S C, HUANG S Y, et al.. A oxygen content magnetomotive analyzer[J]. Opt. Precision Eng., 2007, 15(5):761-766. (in Chinese)[2] GAGNON D, IESSARD S, VERHAEGEN M, et al.. Multiband sensor using thick holographic gratings for sulfur detection by laser-induced breakdown spectroscopy[J]. Applied Optics, 2012, 51(8):7-12.[3] DAVIS M A, KAZMI S M S, PONTICORVO A, et al.. Dependence of vascular fluorescence imaging[J]. Biomedical Optics Express, 2011, 2(12):3349-3362.[4] BORISOV S M, SEIFNER R, KLIMANT I, A novel planar optical sensor for simultaneous monitoring of oxygen, carbon dioxide, pH and temperature[J]. Analytical And Bioanalytical Chemistry, 2012, 403(6):1771-1771.[5] SIEBER A, ENOKSSON P, KROZER A. Smart electrochemical oxygen sensor for personal protective equipment[J]. Sensors Jurnal, IEEE, 2012, 12(6): 1846-1852.[6] 王忠东,马庆万. 土壤中石油矿物油的荧光特性测量[J]. 光学 精密工程, 2010, 18(4):842-847. WANG ZH D, MA Q W. Fluorescence characteristics of mineral oil in soils[J]. Opt. Precision Eng., 2010, 18(4):842-847. (in Chinese)[7] 杨兴华,彭利荣,滕平平,等. 溶胶-凝胶薄膜修饰的微结构聚合物光纤氟离子传感探头[J]. 光学 精密工程, 2010, 18(1):83-86. YANG X H, PENG L R, TENG P P, et al.. Fluoride sensitive optical probe of microstructured polymer optical fiber modified with sol-gel thin film[J]. Opt. Precision Eng., 2010, 18(1):83-86. (in Chinese)[8] MATTEO A, MARC F, ALBERTO C, Luminescence quenching in self-assembled adducts of 2+ complexes and CdTe nanocrystals[J]. Dalton Transaction, 2011, 40(45):12083-12088.[9] VERA L F, GIUSEPPE F, GABRIELLA R. Hyper-efficient quenching of non-conjugated pendant polymer by silver nanoparticles: A demonstration and versatile mechanistic proposition[J]. Chemical Physics Letters, 2012, 532:53-56.