Optimization and validation of the differential interferometric surface plasmon resonance sensor
The sensitivity and dynamic range are main performance parameters of a Surface Plasmon Resonance(SPR) biosensor with phase modulation. In this paper
the main effect factors on sensitivity and dynamic range of the sensor were analyzed
and the influence of nonlinear change of phase difference on the results of real-time monitoring biological reaction was analyzed. A high resolution SPR system based on a Mach-Zehnder configuration was set up. The phase difference curves of a series gold membrane thicknesses and incident angles were simulated by matlab software
and the effect of several factors on the resolution and dynamic range were evaluated. In addition
the real-time monitoring of binding reaction between Bovine Serum Albumin(BSA) and BSA antibodies was also demonstrated. Obtainecl results show that the influence of membrane thickness is significant and nonlinear
but the incident angle shows little effect on the resolution and dynamic range. The influence of narrow dynamic range on biology reaction measurement can be minimized by optimizing some parameters
such as membrane thickness
incident angle and reactants concentration. Experimental results show tha t the sensitivity and dynamic range can be optimized by adjusting gold membrane thicknesses. This paper analyzed several influence factors of the sensitivity and dynamic range of phase modulation SPR biosensors. For a light source with 633 nm wavelength
the most optimal membrane thickness is 48 nm when the reaction between BSA and its antibody is measured. In this situation
the dynamic range is 0.013 6RIU and the sensitivity is 6.6710-7RIU/0.01.
关键词
Keywords
references
PARK K, AHN J, YI S Y, et al.. SPR imaging-based monitoring of caspase-3 activation [J]. Biochemical and Biophysical Research Communications, 2008, 368: 684-689.[2]MORI T, INAMORI K, INOUE Y, et al.. Evaluation of protein kinase activities of cell lysates using peptide microarrays based on surface plasmon resonance imaging[J]. Analytical Biochemistry, 2008, 375: 223-231.[3]SENEVIRATNE A M P B, BURROUGHS M, GIRALT F, et al.. Direct-reversible binding of small molecules to G protein subunits[J]. Biochimica et Biophysica Acta 2011, 1814 : 1210-1218.[4]HIRAGUN T, YANASE Y, KOSE K, et al.. Surface plasmon resonance-biosensor detects the diversity of responses against epidermal growth factor in various carcinoma cell lines [J]. Biosensors and Bioelectronics 2012, 32: 202-207.[5]LIU X, SONG D Q, ZHANG Q L, et al.. Wavelength-modulation surface plasmon resonance sensor [J]. Trends in Analytical Chemistry, 2005, 24(10): 887-893.[6]YANASE Y, HIRAGUN T, KANEKO S, et al.. Detection of refractive index changes in individual living cells by means of surface plasmon resonance imaging [J]. Biosensors and Bioelectronics 2010, 26: 674-681.[7]WONG C L, HO H P, SUEN Y K, et al.. Real-time protein biosensor arrays based on surface plasmon resonance differential phase imaging [J]. Biosensors and Bioelectronics 2008, 24: 606-612.[8]WU S Y, HO H P, LAW W C, et al.. Highly sensitive differential phase-sensitive surface plasmon resonance biosensor based on the Mach-Zehnder configuration[J]. Optics letters, 2004, 29(20): 2378-2380. [9]NARAOKA R, KAJIKAWA K. Phase detection of surface plasmon resonance using rotating analyzer method [J]. Sensors and Actuators B, 2005, 107: 952-956.[10]HO H P, LAW W C, WU S Y, et al.. Phase-sensitive surface plasmon resonance biosensor using the photoelastic modulation technique [J]. Sensors and Actuators B, 2006, 114: 80-84.[11]熊尚,罗雪丰,韩立.纯金膜表面等离子增强的旋光效应[J]. 光学 精密工程,2012,20(7):1525-1531.XIONG SH, LUO X F, HAN L. Plasmon enhanced magneto-optical effect on surface of pure gold film[J]. Opt.Precision Eng., 2012,20(7):1525-1531. (in Chinese)[12]郝鹏,吴一辉.基于噪声分析的波长表面等离子体共振分析仪的数据处理[J]. 光学 精密工程, 2009,17(9):2159-2164.HAO P, WU Y H. Data process of spectroscopic surface plasmon resonance analyzer based on noise analysis[J]. Opt. Precision Eng., 2009,17(9):2159-2164. (in Chinese)[13]马科斯波恩,埃米尔沃耳夫.光学原理,光的传播、干涉和衍射的电磁理论\[M\]. 北京:科学出版社,1978.BORN M, WOLF E. Principles of optics,Electromagnetic theory of propagation, Interference and diffraction of light\[M\]. Beijing:Science press,1978.(in Chinese)[14]SIU P N, CHI M L W, SHU Y W, et al.. Differential spectral phase interferometry for wide dynamic range surface plasmon resonance biosensing [J]. Biosensors and Bioelectronics, 2010, 26: 1593-1598.