ZHAO Hui-jie LIU Xiao-Kang ZHANG Ying. CCD imaging electrical system of AOTF imaging spectrometer[J]. Editorial Office of Optics and Precision Engineering, 2013,21(5): 1291-1296
ZHAO Hui-jie LIU Xiao-Kang ZHANG Ying. CCD imaging electrical system of AOTF imaging spectrometer[J]. Editorial Office of Optics and Precision Engineering, 2013,21(5): 1291-1296 DOI: 10.3788/OPE.20132105.1291.
CCD imaging electrical system of AOTF imaging spectrometer
In consideration of the requirement of Acousto Optics Tunable Filter(AOTF) imagers
a CCD imaging electronic system was designed. The CCD chip CCD57-10 from e2v company was chosen as the image sensor
and the DC/DC+LDO structure was advanced to generate each bias voltage. Then a Field Programmable Gate Array(FPGA)was used to generate driving clock
the AD9826 was taken to convert the CCD output analog signal to a digital signal
and USB or CameraLink interface was adopted to communicate with a computer. Furthermore
a CCD protect circuit was designed and the layout of Printed Circuit Board( PCB) of mixed signals was optimized. The test results for functional units indicate that the system can offer the peak to peak ripple noise of bias voltage less than 10 mV and a AD convert precision of 12 bit. Moreover
the system can image correctly with a image transmit rate of 10 framd/s and the signal to noise ratio better than 54 dB. The system power consumption is less than 5 W. The design can meet the demands of AOTF imaging spectrometers for CCD imaging electronic systems.
关键词
Keywords
references
刘石神.声光可调谐滤波器及其在成像光谱仪上的应用 \[J\]. 红外,2004(7):12-17.LIU SH SH. Acousto-Optic tunable filter and its application in imaging spectrometer \[J\]. infrared, 2004 (7):12-17. (in Chinese)[2]马庆军,宋克非,曲艺.紫外临边成像光谱仪CCD电路系统的设计 \[J\]. 光学 精密工程,2011,19(7):1538-1545.MA Q J, SONG K F, QU Y. Design of CCD circuit systems for ultraviolet limb imaging spectrometers \[J\]. Opt. Precision Eng.,2011,19(7):1538-1545.(in Chinese)[3]王庆有.图像传感器应用技术 \[M\].北京:电子工业出版社,2006.WANG Q Y. Application Technology of Image Sensor \[M\].Beijing:Publishing House of Electronics Industry,2006.(in Chinese)[4]刘金国,余达,周怀得,等.面阵CCD芯片KAI-1010M的高速驱动系统设计 \[J\]. 光学 精密工程,2008,16(9):1622-1628.LIU J G, YU D, ZHOU H D, et al.. High speed driving system design of array CCD chip KAI-1010M \[J\]. Opt. Precision Eng.,2008,16(9) : 1622-1628.(in Chinese)[5]盛翠霞,张涛,纪晶,等.高分辨率CCD芯片FTF4052M的驱动系统设计 \[J\]. 光学 精密工程,2007,15(4):564-569.SHENG C X,ZHANG T,JI J,et al.. Design of CCD FTF4052M driver system with high sensitivity \[J\].Opt. Precision Eng.,2007,15(4) : 564-569. (in Chinese)[6]薛旭成,李云飞,郭永飞.CCD成像系统中模拟前端设计 \[J\]. 光学 精密工程,2007,15(8):1191-1195.XUE X CH, LI Y F, GUO Y F. Design of analog front end of CCD imaging system \[J\].Opt. Precision Eng.,2007,15(8): 1191-1195. (in Chinese)[7]Emi Miyata, Chikara Natsukari, Daisuke Akutsu, et al.. Fast and fexible CCD-driver system using fast DAC and FPGA \[J\]. Nuclear Instruments and Methods in Physics Research, 2001, A 459: 157-164.[8]Ralf Kohley, Juan Manuel Martn-Fleitasa, Lluis Cavaller-Marqus, et al.. CCD camera and data acquisition system of scientific instrument ELMER for the GTC 10-m telescope \[J\]. SPIE, 2004, 5492: 475-483.[9]王辰,王宏强,陈明华,等.星敏感器成像电路噪声分析及降噪处理 \[J\]. 激光与红外,2008,38(7):692-696.WANG C, WANG H Q, CHENG M H, et al.. The Analysis and Disposal of Imaging Circuit Noise Effect on Star Sensor[J]. Laser & infrared, 2008,38(7) : 692-696.(in Chinese)[10]陈伟,黄秋元,周鹏.高速电路信号完整性分析与设计 \[M\].北京:电子工业出版社,2009.CHEN W, HANG Q Y, ZHOU P. Signal Integrity Analysis and Design of High Speed Circuit \[M\]. Beijing: Publishing House of Electronics Industry, 2009. (in Chinese)