. 3-D Finite Element Simulation for measurement of micro-torque of dental implant by SAW devices[J]. Editorial Office of Optics and Precision Engineering, 2013,21(7): 1713-1718
. 3-D Finite Element Simulation for measurement of micro-torque of dental implant by SAW devices[J]. Editorial Office of Optics and Precision Engineering, 2013,21(7): 1713-1718 DOI: 10.3788/OPE.20132107.1713.
3-D Finite Element Simulation for measurement of micro-torque of dental implant by SAW devices
To achieve a non-destructive measurement of stability of a dental implant accurately
a three-dimensional finite element analysis method was used to explore the feasibility to measure the stability of the dental implant by Surface Acoustic Wave(SAW) devices. First
a theoretical model related to the resonant frequencies of SAW devices and the torque under a shear stress gradient field were researched. And then
harmonic responses about the changes of resonant frequencies of the SAW devices under the gradient shear stress field were analyzed by using ANSYS finite element software. Finally
the theoretical model and simulation results were compared. The results show that the resonant frequencies of the SAW devices are in proportion with the maximum shear stress
though there is a 3% deviation between the theory and the simulation results. Also
there is a linear relationship between the torque and the resonance frequencies of the SAW devices. Therefore
the measurement of micro torque of the implant is theoretically feasible by SAW devices
consequently the stability detection of the dental implant can be achieved.
关键词
Keywords
references
LUKAS C, ANTONIN S, RADOVAN S. Influence of the orientation of the osstell transducer during measurement of dental implant stability using resonance frequency analysis:A numerical pooroach[J]. Medical Engineering&Physics,2009,31(7): 764-769.[2]FRIBERG B, SENNERBY L, LINDEN B. Stability measurements of one-stage Branemark implants during healing in mandibles [J]. International Journal of Oral & Maxillofacial Surgery,1999,28 (4): 266-272.[3]QUESADA M P, PRADOS S E, OLMEDO G M V. Measurement of dental implant stability by resonance frequency analysis: A review of the literature [J]. Med. Oral. Patol Oral Cir Bucal, 2009,14 (10) :538-546.[4]GODFREDSEN K,BERGLUNDH T,LINDHE J,et al.. Anchorage of titanium implants with different surface characteristics:an experimental study in rabbits [J]. Clin. Impl. Dent. Relat Res., 2000,2 (3) :120-128.[5]LEE J H, YANG J H. A study on the correlation between implant stability values and initial insertion torque [J]. Journal of Korean Academy of Prosthodontics,2006,44 (3) :314-324.[6]韩韬,施文康.声表面波器件在无线传感系统中的应用[J]. 光学 精密工程,2001,9(6):576-579.HAN T,SHI W K. Applications of the SAW devices in a wireless sensor system [J].Opt. Precision Eng., 2001,9(6): 576-579.(in chinese)[7]BENETTI M, CANNATA D, PIETRANTONIO F D. Pressure sensor based on surface acoustic wave resonators\[C\].Sensors, IEEE Conference,2008 :1024-1027.[8]ANISIMKIN V I, FEDOSOV V I, KRYSTAL R G. Integrated array of SAW sensors for gas analysis\[C\]. Proceedings.IEEE,Sendai:Ultrasonics Symposium,1998:529-533.[9]LIN CH J, HUANG CH W, LIN H P.A study of wireless torque sensing based on SAW sensors \[C\]. Computer Communication Control and Automation, Tainan,2010:211-214.[10]CHEN L, XU CH G, HAO J. Study on passive torque sensor based on SAW resonator [J]. Transducer and Microsystem Technologies,2008,27 (10) :73-75.[11]ZHANG J F, LI Y G, HUI CH. Design and fabrication of floating electrode unidirectional SAW actuator [J].Opt. Precision Eng.,2009, 17(6):1300-1305.[12]ABDOLLAHI A, ZHONG W J. Evaluation on mass sensitivity of SAW sensors for different piezoelectric materials using finite element analysis[J]. IEEE Trans., Ferroelectr,2007,54 :2446-2455.