DUAN Juan, HUA Jian-wen, WANG Zhan-hu etc. Influence of gravitation on moving mirror system in Fourier transform spectrometer[J]. Editorial Office of Optics and Precision Engineering, 2015,23(9): 2613-2619
DUAN Juan, HUA Jian-wen, WANG Zhan-hu etc. Influence of gravitation on moving mirror system in Fourier transform spectrometer[J]. Editorial Office of Optics and Precision Engineering, 2015,23(9): 2613-2619 DOI: 10.3788/OPE.20152309.2613.
Influence of gravitation on moving mirror system in Fourier transform spectrometer
The effect of space weightlessness on the moving mirror system in a Fourier transform spectrometer(FTS) was analyzed. Aiming at two critical factors of gravitation influence on the moving mirror system
solution schemes were put forward
and a validation method in the laboratory was designed. By taking different ground and space environments into account
the influences of gravitation on the moving mirror initial position without a driver and the restoring moment of moving mirror frame with the driver were analyzed. The initial accurate mechanical limit that allows moving mirror motion on a fixed starting point was designed. Then
a control strategy in combination with closed loop PID control and adjustable expected motion law was used to reduce the influence of restoring moment of moving mirror frame. Finally
the moving mirror mechanism of FTS was placed in up-down mode for ground validation. To obtain the bias of positive displacement and negative displacement of the moving mirror and the speed vibration in the uniform area
a test system was established. Experiments show that the bias of positive displacement and negative displacement for the moving mirror motion is less than 6 μm
and the speed vibration(RMS values) tested on upright and up-down frames are respectively 1.4% and 1.1%. The design satisfies the performance of FTS in simpleness and reliability
and provides a application technology for moving mirror systems in space.
关键词
Keywords
references
翁诗甫. 傅里叶变换红外光谱分析[M]. 北京:化学工业出版社,2010. WENG SH F. Fourier Translation infrared spectroscopy [M]. Beijing:Chemical Industry Press,2010. (in Chinese)
李相贤,徐亮,高闽光,等. 分析温室气体及CO2碳同位素比值的傅里叶变换红外光谱仪[J]. 光学 精密工程,2014,22(9):2359-2368. LI X X,XU L,GAO M G,et al..Fourier transform infrared greenhouse analyzer for gases and carbon isotope ratio [J].Opt. Precision Eng.,2014,22(9):2359-2368.(in Chinese)
ANTONIA G, CHRISTOPHER B, WALTER W, et al..An experiment using high spectral resolution CrIS measurements for atmospheric trace gases:carbon monoxide retrieval impact study [J]. IEEE Geoscience and Remote Sensing Letters, 2014,11(9):1639-1643.
BLUMSTEIN D, CHALON G, CARLIER T, et al.. IASI instrument:technical overview and measured performances [J]. Infrared Spaceborne Remote Sensing XII, Proceedings of SPIE, 2004,5543:196-207.
MARC-ANDRE S, FRANCOIS C, CHRISTOPHE D, et al.. ACE-FTS instrument detailed design [J]. Earth Observing Systems VII, Proceedings of SPIE,2002,4814:70-81.
RONALD J G, FREDERICK L W, NATHAN F. Cross-track Infrared Sounder (CrIS) development status [J]. Infrared Spaceborne Remote Sensing XI, Proceedings of SPIE, 2003, 5152:1-8.
BICKNELL W E. Overview of GOES High-resolution interferometer sounder(GHIS) brassboard development and test program [J]. GOES-8 and Beyond, Proceedings of SPIE, 1996, 2812:389-395.
RICHARD J K, SCOTT D L. Mechanical design of the crosstrack infrared sounder (CrIS) [J]. Infrared Spaceborne Remote Sensing IX, Proceedings of SPIE,2002, 4486:445-455.
齐乃明,张文辉,高九州,等. 空间微重力环境地面模拟试验方法综述 [J]. 航天控制,2011,29(3):95-100. QI N M,ZHANG W H,GAO J ZH, et al..The primary discussion for the ground simulation system of spatial microgravity [J]. Aerospace Control, 2011,29(3):95-100.(in Chinese)
夏翔. 干涉光谱仪多速度模式扫描技术研究[D]. 上海:中国科学院上海技术物理研究所,2014. XIA X.Research on technology of muti-speed mode scanning of interference spectrometer [D].Shanghai:Shanghai Institute of Technical Physics, Chinese Academy of Sciences,2014. (in Chinese)
孙麟治,李鸣鸣,程维明. 精密定位技术研究[J]. 光学 精密工程,2005,13(1):69-75. SUN L ZH,LI M M,CHENG W M. Study on precision positioning technique [J].Opt. Precision Eng.,2005,13(1):69-75.(in Chinese)
胡寿松. 自动控制原理[M]. 北京:科学出版社,2013. HU SH S.The Principle of Automatic Control [M]. Beijing:Science Press, 2013. (in Chinese)
许忠仁,姜丽,穆克,等. 基于FPGA的数字PID控制器研究[J]. 辽宁石油化工大学学报,2012,32(3):74-79. XU ZH R, JIANG L,MU K, et al..Digital PID controller based on FPGA [J].Journal of Liaoning Shihua University,2012,32(3):74-79. (in Chinese)
谢元安, 韩志刚. 基于白光干涉的空间频域算法研究[J]. 光电工程,2011,38(7):81-85. XIE Y A,HAN ZH G. Study of spatial frequency domain analysis method based on white light interferometry [J]. Opto-Electronic Engineering, 2011,38(7):81-85. (in Chinese)