Linear calibration for on-machine measurement of laser probe pose
A calibration method suitable for a laser probe in on-machine laser scanning measurement was proposed to eliminate the measuring errors from installation pose. A prototype system for the on-machine laser scanning measurement was built and a model of the laser probe moved with a machine tool was established. The center of a standard ball was fitted by scanning the ball surface with multi-perspective
and a linear algorithm to solve the installation pose parameters of the laser probe was given to overcome the heavy computation and un-robustness of the existing nonlinear algorithms. The influence of the machine tool movement error on the measurement was also analyzed and an error model was established and a compensation method for the machine tool error was proposed. The experimental results show when the standard ball with a known diameter was measured
its diameter error measured by the laser probe at different angles is less than 0.05 mm
and the position error of the standard ball center is reduced by 83% with the compensation method. The results illustrate the feasibility of the calibration method as well as the validity of the error model and the compensation method.
关键词
Keywords
references
罗胜彬, 宋春华, 韦兴平, 等. 非接触测量技术发展研究综述[J]. 机床与液压, 2013, 41(23):150-153. LUO SH B, SONG CH H, WEI X P, et al.. Review for the development and research of non-contact measurement technology[J]. Machine Tool & Hydraulics, 2013, 41(23):150-153. (in Chinese)
裘祖荣, 石照耀, 李岩. 机械制造领域测量技术的发展研究[J]. 机械工程学报, 2010, 46(14):1-11. QIU Z R, SHI ZH Y, LI Y. Research on the development of measurement technology mechanical manufacture[J]. Journal of Mechanical Engineering, 2010, 46(14):1-11. (in Chinese)
CHUN B S, KIM K, GWEON D. Three-dimensional surface profile measurement using a beam scanning chromatic confocal microscope[J]. Review of Scientific Instruments, 2009, 80(7):30-34.
ONG W H, CHIU W K. Non-contact measurement techniques for structural health monitoring[J]. Procedia Engineering, 2014, 75:45-50.
王龙君, 王亚平. 基于五轴数控机床的激光在线检测方法研究[J]. 计算机测量与控制, 2008, 16(3):297-298, 326. WANG L J, WANG Y P. Study on laser on-line detection based on 5-axis CNC machine tools[J]. Computer Measurement & Control, 2008, 16(3):297-298, 326. (in Chinese)
孙玉文, 郭东明, 贾振元. 复杂曲面的测量加工一体化[J]. 科学通报, 2015, 60(9):781-791. SUN Y W, GUO D M, JIA ZH Y. Integrated measurement and machining technology of complex surfaces[J]. Chinese Science Bulletin, 2015, 60(9):781-791. (in Chinese)
房佳威, 张国雄, 裘祖荣. 非接触激光测头测量轴线标定系统[J]. 电子测量技术, 2006, 29(3):160-161. FANG J W, ZHANG G X, QIU Z R. Calibration system for surveying axis of non-contact laser probe[J]. Electronic Measurement Technology, 2006, 29(3):160-161. (in Chinese)
卢科青, 王文, 陈子辰. 点激光测头激光束方向标定[J]. 光学精密工程, 2010, 18(4):880-886. LU K Q, WANG W, CHEN Z CH. Calibration of laser beam-direction for point laser sensors[J]. Opt. Precision Eng., 2010, 18(4):880-886. (in Chinese)
BI C, LIU Y, FANG J G, et al.. Calibration of laser beam direction for optical coordinate measuring system[J]. Journal of the International Measurement Confederation, 2015, 73:191-199.
刘书桂, 陈超. 基于改进遗传算法的三坐标测量机光学测头的标定[J]. 天津大学学报, 2005, 38(8):695-700. LIU SH G, CHEN CH. Calibration of CMM optical probe based on improved genetic algorithm[J]. Journal of Tianjin University, 2005, 38(8):695-700. (in Chinese)
费致根, 郭俊杰, 马小军, 等. 微型五坐标测量机激光测头安装误差的标定[J]. 西安交通大学学报, 2011, 45(5):64-68. FEI ZH G, GUO J J, MA X J, et al.. Installation error calibration of laser probe for miniature five-coordinate measuring machine[J]. Journal of Xi'an Jiaotong University, 2011, 45(5):64-68. (in Chinese)