Hong ZHONG, Wei TAO, Hui ZHAO, et al. Measurement of magnetic bead concentration based on laser transmission method[J]. Optics and precision engineering, 2017, 25(2): 297-303.
DOI:
Hong ZHONG, Wei TAO, Hui ZHAO, et al. Measurement of magnetic bead concentration based on laser transmission method[J]. Optics and precision engineering, 2017, 25(2): 297-303. DOI: 10.3788/OPE.20172402.0297.
Measurement of magnetic bead concentration based on laser transmission method
Magnetic beads are widely used in biological experiments
such as immunoassay and cell separation. The detection of magnetic concentration is profound to the evaluation of experimental results. Therefore a magnetic beads concentration measuring system with dual optical path based on laser transmission method was presented. The concentration of magnetic beads was determined by measuring the ratio of the intensity of incident laser to the intensity of the transmitted laser. A series of magnetic bead suspensions with various concentrations were tested. For each concentration
the measurement was performed for many times and a total of 8
000 sets of data were collected. By fitting the measured data
the relational function between beads concentration and output data can be obtained. The relation between the concentration of magnetic beads and output data was well fitted exponentially. The fitting residual sum of squares is 0.000 088 89 and the determination coefficient is 0.997 1
which illustrate good fitting ability of the model. The experimental results show that the relative measurement error is less than 2.5% and the output data fluctuates within 2.5%
which indicate that the beads concentration measurement system based on laser transmission has high measurement accuracy and good repeatability.
WEN Y M, LI ZH Q, TONG J Y, et al.. Rapid detection of Listeria monocytogenes by immunomagnetic separation combined with selective medium[J]. Chin. J. Biotech., 2013, 29(5):672-680.(in Chinese)
GUO H F, ZHANG W H, WEN D Q, et al.. A new method to detect antibodies based on reusable immunomagnetic beads[J]. Immunological Journal, 2006, 22(5):577-579. (in Chinese)
ZHANG X L.A method of detecting magnetic beads concentration based on low field NMR and its application:Chinese patent, 201410326418.1[P].2014-10-29. (in Chinese)
ZHENG Y S, JACQUEMOD C, SAWAN M. A portable lab-on-chip platform for magnetic beads density measuring[C].2013 IEEE International Symposium on Circuits and Systems (ISCAS2013), IEEE, 2013:1071-1074.
LIU R P, LIU Q, QI ZH M.Interference-resistant turbidity detector based on measurement of scattered light power ratio[J]. Opt. Precision Eng., 2011, 19(6):1221-1227.(in Chinese)
MIZAIKOFF B. Infrared optical sensors for water quality monitoring[J]. Water Science and Technology, 2003, 47(2):35-42.
MENG Q L, SUN G Q, TONG G X. A measuring apparatus for turbidity of liquids[J]. Chinese Journal of Scientific Instruments, 2005, 26(8):252-253.(in Chinese)
YU Q, YU X F, CUI CH C.Design of laser parallel confocal measurement system with single source and dual beam paths[J].Opt. Precision Eng., 2013, 21(2):281-285.(in Chinese)
LI Y P, ZHANG G J, LI Q B. Infrared CO 2 gas sensor based on space double beams and its measurement model[J].Opt. Precision Eng., 2009, 17(1):14-19. (in Chinese)
WANG Q Q, LIU J, PENG ZH, et al..Measurement system for laser divergence angle based on LabView[J].Chinese Journal of Lasers, 2012, 39(11):1108005. (in Chinese)