Jing LI, Jian-ke ZHAO, Ming CHANG, et al. Radiometric calibration of photographic camera with a composite plane array CCD in laboratory[J]. Editorial office of optics and precision engineeri, 2017, 25(1): 73-83.
DOI:
Jing LI, Jian-ke ZHAO, Ming CHANG, et al. Radiometric calibration of photographic camera with a composite plane array CCD in laboratory[J]. Editorial office of optics and precision engineeri, 2017, 25(1): 73-83. DOI: 10.3788/OPE.20172501.0073.
Radiometric calibration of photographic camera with a composite plane array CCD in laboratory
To eliminate the serious vignetting phenomenon and to solve the difficulty of choosing a relatively positioning target value problems among CCDs during the radiometric calibration of the whole image plane of the photographic camera with a composite plane array CCD
a method for radiometric calibration of multi-CCDs was proposed. After the dark signal calibration on each pixel
the gray value of the pixel in the vignetting area was revised
and by choosing the appropriate relative calibration target gray value of the whole image plane
the radiometric calibration of the photographic camera with a composite plane array CCD was eventually completed. Through analysis of the gray distribution characteristics of vignetting area
the gray correction method for all kinds of explosion time and radiance was proposed. With calculation of fitting coefficient between each CCD gray value and the input radiance respectively
the coefficient with global minimum fitting error was chosen to calculate the relative calibration target gray value corresponding to radiance. After the radiometric calibration using this method
the non-uniformity of the whole image plane of the photographic camera with a composite plane array CCD is reduced from above 20% to better than 2%
and the accuracy of the absolute calibration is 4.23%. The result indicates that the proposed calibration method is appropriate for the calibration of composite planes array CCD. The accuracy of the calibration satisfies the radiometric calibration requirement of aerial cameras.
REN H H, RUAN P, HE J W, et al.. Study of the radiometric calibration of TDI-CCD spatial stereo camera[J]. Acta Optica Sinica, 2010,30(12):3476-3480. (in Chinese)
TAO M H. Research on the Vignetting Correction Method of Mosaic Images Based on Reflector[D]. Changchun:Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, 2012.(in Chinese)
BOIVIN L P. Radiometric errors caused by diffraction from circular apertures:edge effects[J]. App. Opt., 1977, 16(2):377-384.
乔闹生. CCD非线性及其校正研究[J]. 光子学报,2008,37(11):2305-2309.
QIAO N SH. Nonlinearity of CCD and its adjustment[J]. Acta Photonica Sinica, 2008,37(11):2305-2309.(in Chinese)
LI X SH, YE ZH, REN J W, et al.. Processing method for abnormal response of space camera image in orbit[J]. Chinese Journal of Liquid Crystals and Displays, 2012,27(4):557-562. (in Chinese)
SUN J X. Research on Radiometric Calibration for Large-aperture Space Optical Remote Sensor[D]. Changchun:Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, 2015.(in Chinese)
郑亮亮,金光,曲宏松,等. 高信噪比星载CCD成像电路系统[J]. 光学精密工程, 2016,24(8):2027-2036 ZHENG L L, JIN G, QU H S, et al.. Space-borne CCD imaging circuit system with high signal-to-noise ratio[J].Opt. Precision Eng., 2016, 24(8):2027-2036.(in Chinese)
ZHAO W N, HU X Q, FANG W, et al.. Development and applications of intercalibration for satellite optical instruments[J].Opt. Precision Eng., 2015, 23(7):1921-1931.(in Chinese)
XIU J H, HUANG P, LI J, et al.. Radiometric calibration of large area array color CCD aerial mapping camera[J].Opt. Precision Eng., 2012,20(6):1368-1373. (in Chinese)
LIU H X, LI X SH, LI X J, et al..Correction of vignetting image for reflector-based mosaic CCD camera[J]. Opto-Eletronic Engineering, 2015,42(2):9-14. (in Chinese)
EMVA Standard 1288. Standard for characterization of image sensors and cameras[S].European Machine Vision Association,2012.