Hong-xing LIU, Xian-sheng LI, Jing-xu SUN, et al. Haze and transmittance real-time tester for curved-surface and transparent samples[J]. Optics and precision engineering, 2018, 26(12): 2881-2887.
DOI:
Hong-xing LIU, Xian-sheng LI, Jing-xu SUN, et al. Haze and transmittance real-time tester for curved-surface and transparent samples[J]. Optics and precision engineering, 2018, 26(12): 2881-2887. DOI: 10.3788/OPE.20182612.2881.
Haze and transmittance real-time tester for curved-surface and transparent samples
In order to realize the real-time measurement of haze and light transmittance for large-surface transparent samples
a haze and transmittance tester using LED as the light source and a double integrating sphere as the signal measurement part was designed. First
the theoretical derivation of the light transmittance and haze was completed on the basis of the system scheme. Second
simulation analysis of the light source spectral distribution affecting measurement
optical design
integrating sphere parameter design
and support adjustment mechanism design were completed
increasing the test accuracy and efficiency. Finally
calibration and sample testing of the haze and transmittance real-time tester were conducted. The test results show that the transmittance absolute error of the standard haze sample is less than 0.4% and its haze absolute error is less than 0.3%. Moreover
the transmittance deviation between the measured value and theoretically calculated value of the samples from the subscriber is also less than 0.4%
which confirms the claim of the subscriber.
关键词
Keywords
references
卢进军, 刘卫国, 潘永强编著.光学薄膜技术[M].北京:电子工业出版社, 2011.
LU J J, LIU W G, PAN Y Q. Optical Thin Film Technology [M].Beijing:Electronic Industry Press, 2011.(in Chinese)
National standard of the People's Republic of China GB/T2410-2008.Determination of the luminous transmittance and transparent plastics[S].Beijing: Standards Press of China, 2008.(in Chinese)
Honeywell International. Mesearment equipment and methods for flaky material and other materials: Chinese Patant, CN 102422148 A[P].2012-04-18.(in Chinese)
CHEN F, ZHENG X B. Influence of spectrum not-matching on calibration precision of remote sensor [J]. Opt. Precision Eng., 2008, 16(3): 415-419. (in Chinese)
ZHAO X, WU H CH, SONG L P, et al.. Design of freeform LED concentrator for micro-project [J]. Opt. Precision Eng., 2010, 18(10): 2123-2129. (in Chinese)
LIH H X, REN J W, LIU Z X, et al.. Design of integrating sphere solar spectrum simultor based on xenon lamp and LEDs [J]. Opt. Precision Eng., 2012, 20(7): 1447-1456. (in Chinese)