A non-contact respiratory and heart rate monitoring system was designed to monitor bedridden patients. Initially
according to the mechanical properties of the cardiac ejection contraction process
piezoelectric ceramic sensors with high sensitivity and good stability were selected to acquire ballistocardiography (BCG) signals. The collected signal was then desiccated
filtered and amplified; it was then digitized to obtain a heart impact diagram. A respiratory signal was then extracted from the cardiogram with smooth filtering
and the respiratory rate signal was obtained via an FFT transformation. The respiratory envelope and high-frequency interference of the BCG signal were removed using a band-pass filter
and the j-wave peak measurement of the BCG signal was obtained to calculate the heart rate. Finally
to verify the accuracy and consistency of the system
it was compared with respiratory data collected by BIOPAC and ECG signals. The respiratory error rate of the system is less than 4.5% and the heart rate error is 9.7%. A Bland-Altman analysis indicates that the heart rate measurement and accuracy computed by the monitoring system are consistent with those of BIOPAC.
CAI J. Application of bedside responsibility nursing in cardiovascular medicine and effect on sleep quality[J]. World Journal of Sleep Medicine , 2018, 5(6): 715-706. (in Chinese)
LAKSHMI D B, VISHVAKSENAN K S, SENTHAMIL S K, et al.. Patient monitoring system using cognitive internet of things[J]. Journal of Medical Systems , 2018, 42(11): 229.
ZHANG W P, KUMAR M, YU J F, et al.. Medical long-distance monitoring system based on internet of things[J]. Eurasip Journal on Wireless Communications & Networking , 2018, 1: 176.
WEENK M, VAN G H, FRIETMAN B, et al.. Continuous monitoring of vital signs using wearable devices on the general ward: pilot study[J]. Jmir Mhealth & Uhealth , 2017, 5(7): e91.
GANESAN M, LAVANYA R, SUMESH E P. A survey on ballistocardiogram to study the mechanical activity of heart[C]. 2017 International Conference on Communication and Signal Processing ( ICCSP ), Chennai, India: IEEE , 2017: 557-561.
NEDOMA J, KEPAK S, FAJKUS M, et al.. Magnetic resonance imaging compatible non-invasive fibre-optic sensors based on the bragg gratings and interferometers in the application of monitoring heart and respiration rate of the human body: a comparative study[J]. Sensors , 2018, 18(11): 3713.
WUSK G, GABLER H. Non-invasive detection of respiration and heart rate with a vehicle seat sensor[J]. Sensors , 2018, 18(5): 1463.
BRUSER C, ZINK M.D.H, WINTER S, et al.. A feasibility study on the automatic detection of atrial fibrillation using an unobtrusive bed-mounted sensor[C]. 2011 Conference on Computing in Cardiology, Hangzhou, P.R.Chian: IEEE, 2011: 13-16.
ZINK M D, BRUSER C, STUBEN B O, et al.. Unobtrusive nocturnal heartbeat monitoring by a ballistocardiographic sensor in patients with sleep disordered breathing[J]. Scientific Reports , 2017, 7(1): 13175.
ETEMADI M, HERSEK S, TSENG J M, et al.. Tracking Clinical status for heart failure patients using ballistocardiography and electrocardiography signal features[C]. 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Chicago, USA: IEEE, 2014: 5188-5191.
NAZIYOK T P, ZELEKE A A, ROHRIG R. Contactless patient monitoring for general wards: a systematic technology review[J]. Studies in Health Technology & Informatics , 2016, 228:707-711.
ZHANG X W, ZHANG L Y, DING L CH, et al.. Heartbeat detection based on ballistocardiograms[J]. Journal of Tsinghua University ( Science and Technology ), 2017, 57(7): 763-767. (in Chinese)
XIAO L, LI H L, ZHANG X W, et al.. A wearable ballistocardiogram-electrocardiogram union acquisition system[J]. Journal of Biomedical Engineering , 2018, 35(5): 727-732. (in Chinese)
DU G F, LI Z, SONG G B. A PVDF-based sensor for internal stress monitoring of a Concrete-Filled Steel Tubular (CFST) column subject to impact loads[J]. Sensors , 2018, 18(6): 1682.
WANG M Y, XU Z ZH, YUE S L, et al.. Calibration of PVDF pressure gauge and its application in dynamic parameters tests of similar materials[J]. Transactions of Beijing Institute of Technology , 2017, 37(10): 991-997. (in Chinese)
JIA D G, ZHAO J, ZHANG H X, et al.. Anti-electromagnetic interference electrocardiogram monitoring system[J]. Opt. Precision Eng. , 2018, 26(5): 1165-1174. (in Chinese)
DAS S, PAL S, MITRA M. Real time heart rate detection from PPG signal in noisy environment[C]. 2016 International Conference on Intelligent Control Power and Instrumentation ( ICICPI ), Kolkata, India: IEEE, 2017: 70-73.
YY 1079-2008, 心电监护仪[S].
YY 1079-2008, Electrocardiographic Monitors[S]. (in Chinese)
LI Y CH, LI X S. Evaluation on different assessment methods of consistency of quantitative measurements[J]. Modern Preventive Medicine , 2007, 34(17): 3263-3266. (in Chinese)
GIAVARINA D. Understanding bland altman analysis[J]. Biochemia Medica , 2015, 25(2): 141-151.
BLAND J M, ALTMAN D G. Statistical methods for assessing agreement between two methods of clinical measurement[J]. The Lancet , 1986, 327(8476): 307-310.