浏览全部资源
扫码关注微信
哈尔滨工业大学 超精密光电仪器工程研究所,黑龙江 哈尔滨,150001
收稿日期:2005-07-19,
修回日期:2005-09-01,
网络出版日期:2005-12-30,
纸质出版日期:2005-12-30
移动端阅览
敖 磊, 谭久彬, 崔继文, 等. 激光CCD二维自准直仪中圆目标中心精确定位算法[J]. 光学精密工程, 2005,13(6):668-673.
AO Lei, TAN Jiu-bin, CUI Ji-wen, et al. Precise center location algorithm for circle target in CCD laser dual axis autocollimator[J]. Optics and precision engineering, 2005, 13(6): 668-673.
提出一种新的激光CCD二维自准直仪中圆目标中心精确定位算法。该算法采用变结构元多尺度广义数学形态学滤波算法滤除图像噪声
由Zernike矩定位算法得出目标轮廓的亚像素位置
用最小二乘拟合法精确获取圆目标的中心位置
并进行了实验。结果表明:该算法稳定性好
定位精度高
应用该算法后自准直仪的测量分辨力提高了10倍
在±2'测量范围内测量不确定度优于0.3″
有效地提高了自准直仪的测量分辨力和测量精度。
A precise circle target center location algorithm was proposed and the multiscale generalized mathematical morphological filtering method with multiple structuring elements was utilized for eliminating noise in the image. The subpixel position of circle target was obtained by Zernike moment contour subpixel location algorithm
and the circle target center position was acquired ultimately and accurately using circle fitting method based on least square method. The experiment results show that the method is stable and accurate located for measuring 2D small angle
and the measuring uncertainty of this modified instrument is better than 0.3″ within ±2' range
with the measuring resolution increased by 10 times. The results also show that the measuring resolution and accuracy of the autocollimator can be improved effectively.
. YANDAYAN T, AKG Z S A, HAITJEMA H. A novel technique for calibration of polygon angles with non-integer subdivision of indexing table[J]. Precision Engineering-Journal of the International Societies for Precision Engineering and Nanotechnology, 2002, 26 (4): 412-424.
. 魏仲慧. 数字光电自准直仪的研究与设计[J]. 光学 精密工程. 1996,4(3):116-120. WEI ZH H. Research and design of the digital optoelectronic autocollimator[J]. Optics and Precision Engineering, 1996,4(3):116-120.(in Chinese)
. KONIAKHINE I A, TIMOFEEV A N, MUSYAKOV V L. The optic-electronic systems for control position of the elements unblocked aperture radio telescope[J]. Proceedings of the Third International Symposium on Instrumentation Science and Technology, 2004,3: 1-5.
. JIE Y, LONG X. CCD-area-based autocollimator for precision small-angle measurement[J]. Review of Scientific Instruments, 2003,74(3): 1362-1365.
. LEVITON D B. Ultrahigh resolution absolute Cartesian electronic autocollimator[J].SPIE, 2003,5190:468-475.
. 张尧禹,张明慧,乔彦峰. 一种高精度CCD激光自准直测量系统的研究[J]. 光电子·激光.2003,14(2):168-170. ZHANG Y Y, ZHANG M H, QIAO Y F. Research of a CCD laser collimation system with high precision[J]. Journal of Optoelectronics·Laser, 2003,14(2):168-170.(in Chinese)
. ZHAO M R,LIN Y CH,ZHANG P,et al. Research on improving the measuring accuracy of laser autocollimator used over long distance[J].SPIE, 2002,4927:300-304.
. CUI J, QIU L, JIN G L, et al. A new method used for measuring the parameter of elastic thin-neck based on CCD aiming . Proceedings of the Second International Symposium on Instrumentation Science and Technology, 2002:799-803.
0
浏览量
660
下载量
18
CSCD
关联资源
相关文章
相关作者
相关机构
京公网安备11010802024621