XIAO Sha-li, SHI Jun, LEI Xiao-ming, et al. Characteristics and application of elliptical curved LiF crystal analyzer[J]. Optics and precision engineering, 2007, 15(6): 824-828.
XIAO Sha-li, SHI Jun, LEI Xiao-ming, et al. Characteristics and application of elliptical curved LiF crystal analyzer[J]. Optics and precision engineering, 2007, 15(6): 824-828.DOI:
an X-ray elliptical curved crystal analyzer whose measurement range
focal length
eccentricity and Bragg angel range are respectively 0.6~6 keV
1350 mm
0.9586 and 30°~60° was designed for investigating the laser producing plasmas. Aimed to the characteristic study of a high density plasma
a experiment was carried out with laser length of 0.35 μm
power of 1.6×10
14
W/cm
2
and the incident angel on the Ti plane of 45°. The result shows that the sensitivity of elliptical curved LiF crystal has very good detection ability
its calculated wavelength resolution is about 500~1 000 in the range of 0.2~0.35 nm. Experimental results also show that the curved crystal analyzer can be used in spatial resolution measurement for aplanatic property
its sensitivity is higher than a flat crystal. It suggests that curved LiF crystal analyzer is suitable for plasma diagnostic.