relative luminance and excitation spectra characteristics are important factors in the field of phosphors research. Measurement of emission spectra is relative simple
its precision and repeatability can be ensured within 1% more easily
but relative luminance and excitation spectra is difficult to have precision and repeatability of 1% affected by test method. In this article
one method that phosphors surface diffuse reflection intensity is measured to improve relative quantum efficiency accuracy of phosphors for PDP in UV-VUV range between 120nm to 300nm
the relative quantum efficiency measuring precision is below 1% and the comparative subtle excitation spectrum curve is obtained
the spectra resolution is 1nm. At last
the measurement of excitation spectrum of (BaMgAl10O17:Eu) blue phosphors and (LaPO4:Tb