a confocal endscopic imaging system is established
its axial and lateral resolution has been measured and analyzed
and its application in pig skin has been investigated. First
based on the requirement of system
principles and configuration of system are presented. Then the theoretical values of spatial resolution are shown in confcoal endoscope. Finally
the system is tested
and the reason for difference between measuring date and theoretical value is illuminated. Experimental results indicate that its axial and lateral resolution is respectively 10 and 1.9
and it can be used to obtained tomographic image in pig skin.