李亚萍, 张广军, 李庆波. 空间双光路红外CO<sub>2</sub>气体传感器及其测量模型[J]. 光学精密工程, 2009,17(1): 14-19
LI Ya-ping, ZHANG Guang-jun, LI Qing-bo. Infrared CO<sub>2</sub> gas sensor based on space double beams and its measurement model[J]. Editorial Office of Optics and Precision Engineering, 2009,17(1): 14-19
提出了一种基于空间双光路结构的红外CO
2
气体传感器。阐述了传感器光学探头的结构和工作原理
并给出了气室的几何形状和相应的尺寸参数。以朗伯-比尔定律为理论依据
结合空间双光路结构的工作机理
以及红外光源周期性开关的工作方式
推导出了该传感器的线性测量模型
并采用最小二乘法对建立的测量模型进行标定。在静态下对该传感器的传感特性进行了测试
测试结果表明:在0~3%的测量范围内
该传感器的实验值与真实值符合较好
其精度为0.024 2%;系统运行良好
可无故障连续在线测量30 d以上。该传感器能有效消除温度信号和背景光信号的干扰
能够在5~40 ℃的温度内、不同杂光背景下正常工作
基本满足CO
2
气体浓度测量的功耗低、精度高、稳定性好等要求。
An infrared CO
2
gas sensor was developed based on space-double-beam structure. First
the structure and working principle of the optical probe were introduced.A new-type gas cell was presented and the optimal size of the gas cell were calculated in detail. Then
the linear mathematical model of the sensor was established based on Lambert-Beer Law
the principle of space-double-beam optical structure and the working model of an infrared light source in turning on and off periodically. Finally
the unknown coefficients in the mathematical model were calculated with Least Squares Method. The characteristics of the infrared sensor were tested in static state
tested results show that the values obtained from the experiment is close to the true values and the accuracy of the sensor is 0.024 2% when measurement range is 0~3%. The system can work continuously more than 30 days without the disturbance from the temperature signal and background optical signal
it also can work at 5~40 ℃ and on different background conditions. These results conclude that the sensor can satisfy the system requirements of lower-power-consume
higher-precision and stability.
张广军,武晓利. 新型高性能红外二氧化碳传感器 . 红外与激光工程,2002,31(6): 540-544. ZHANG G J, WU X L. Novel IR carbon dioxide sensor with high performance [J]. Infrared and Laser Engineering, 2002, 31(6): 540-544. (in Chinese)[2] GIULIANO B, CARLA C, REMO T. Measurement of long-term average carbon dioxide concentration using passive diffusion sampling [J]. Atmospheric Environment, 2004, 38: 1625-1630.[3] ROBERT F, THOMAS T. An automotive bi-source spectroscopic carbon dioxide sensor with pressure compensation [J]. Sensors and Actuators B, 2007, 127: 82-88.[4] CONOR S. Development of an optical sensor probe for the detection of dissolved carbon dioxide [J]. Sensors and Actuators B, 2006, 119: 288-294.[5] 程基伟,张琦. 材料腐蚀预测数学模型的研究[J]. 航空学报,2000,21(2): 183-186. CHENG J W, ZHANG Q. Study of mathematical models in material corrosion prediction [J]. Journal of Aeronautics, 2000, 21(2): 183-186. (in Chinese)[6] 朱坚民,王中宇. 基于神经网络的测量模型的建立及检验[J]. 光学 精密工程,2000,8(4): 9-10. ZHU J M, WANG ZH Y. Establishment and verification of measurement model based on neural networks [J]. Opt. Precision Eng., 2000, 8(4): 9-10. (in Chinese)[7] 康永济.红外气体分析器[M]. 北京:化学工业出版社,1993. KANG Y J. Infrared Gas Analyzer [M]. Beijing: Chemistry Industry Press, 1993. (in Chinese)[8] 张广军,吕俊芳,周秀银,等. 新型红外二氧化碳分析仪[J]. 仪器仪表学报,1997,18(2) :134-137. ZHANG G J, LV J F, ZHOU X Y, et al.. New infrared carbon dioxide gage [J]. Chinese Journal of Scientific Instrument, 1997, 18(2): 134-137. (in Chinese)[9] ZHANG G J, WU X L. A novel CO2 gas analyzer based on IR absorption [J]. Optics and Lasers in Engineering, 2004, 42(2): 219-231.[10] YOSHIHARU T. Fabrication and characterization of an infrared-absorption-type CO2 gas sensor using a PLZToptical shutter [J]. Electronics and Communications in Japan(Part II:Electronics)1999, 82(4): 1-8.[11] 黄富贵,崔长彩. 评定直线度误差的最小二乘法与最小包容区域法精读之比较[J]. 光学 精密工程,2007,15(6): 879-892. HUANG F G, CUI CH C. Comparison of evaluating precision of straightness error between least square method and least envelope zone method [J]. Opt. Precision Eng., 2007, 15(6): 879-892. (in Chinese)[12] 李子芳,崔长彩,车仁生,等. 基于遗传算法的圆度公差评定法与采用最小二乘法评定的比较[J]. 光学 精密工程,2003,11(3): 256-261. LI Z F, CUI CH C, CHE R SH, et al.. Comparison of genetic algorithm based evaluation of roundness with evaluation of roundness based on least squared method [J]. Opt. Precision Eng., 2003, 11(3): 256-261. (in Chinese)[13] 刘俊,秦岚,刘京诚,等. 一种新型差动式压电加速度传感器[J]. 光学 精密工程,2007,15(6) :903-909. LIU J, QIN L, LIU J CH, et al.. A novel differential piezoelectric accelerating sensor [J]. Opt. Precision Eng., 2007,15(6): 903-909. (in Chinese)