LI Na-na, AN Zhi-yong, CUI Ji-cheng. Aligning and testing of optical system of echelle spectrography[J]. Editorial Office of Optics and Precision Engineering, 2009,17(3): 531-536
LI Na-na, AN Zhi-yong, CUI Ji-cheng. Aligning and testing of optical system of echelle spectrography[J]. Editorial Office of Optics and Precision Engineering, 2009,17(3): 531-536DOI:
the characteristics of echelle grating spectrography were introduced and a method of fixing and checking for the optical system of the echelle spectrograph was researched.Proposed method could reach the precision of the design through fixing the slot
collimation lens
prism
echelle grating and focusing lens.A special fixing method for the echelle grating spectrography during the fixing process was also introduced. Finally
a calculation formula was deduced based on Reyleigh principle to check the resolution of the echelle grating spectrography. The checked result shows that the resolution of the echelle grating spectrography meet the system requerement of 10 000.These results reported here mean that propose method can offer a reference for fixing and adjusting of other precision instruments.
关键词
Keywords
references
武旭华,朱永田,王磊. 高分辨率阶梯光栅光谱仪的光学设计[J]. 光学 精密工程,2003,13(5):442-447. WU X H,ZHU Y T,WANG L. Optical design of high resolution echelle spectrograph[J]. Opt. Precision Eng.,2003,13(5):442-447.(in Chinese)[2] HARRION G R,ARCHER J E,CAMUS J.A fixed-focus broad-range echelle spectrograph of high speed and resolving power[J]. J. Opt. Soc. Am.,1952,42(10):706-712.[3] ZHU Y, XU W. Optical and IR telescope instrumentation and detectors . Washington, USA: SPIE, 2000:141-147.[4] WANG SH G, SU D Q,CHU Y Q, et al.. Special configuration of a very large Schmidt telescope for extensive astronomical spectroscopic observation[J]. Applied Optics, 1996, 35(25):5155-5161.[5] TULL R G. Instrumentation in astronomy Ⅷ . Washington, USA: SPIE, 1994:674-684.[6] SCHROEDER D J. An echelle spectrometer-spectrograph for astronomical use[J]. Applied Optics, 1967, 6(11):1976-1980.