A novel nondestructive measuring method for the refractive index of a transparency capillary is introduced in this paper. When a collimated light passes through a capillary filled in a standard liquid with known refractive index
the focal position is dependent mainly on the wall-material of the capillary.On the basis of the above
the position and focal length of a capillary are imaged by a lens onto a CCD chip accurately
when the capillary is filled in two different liquids with known refractive indexes
respectively.Then
the refractive index of the wall- material is obtained by measuring and calculating two positions and focal lengths separately. The measurement accuracy of refractive index is better than 0.005. The effects of measuring parameters
including inner and outer diameters of the capillary
the refractive index of satandard liquid
as well as the image depth and sensitivity of refractive index on the measuring precision are analyzed. It is shown that the thicker capillary wall is
the higher the measuring precision is
and the lower the refractive index of standand liquid is
the higher the calibrating accuracy is. The introduced method is characterized by simple equipment
easy to operation
convenient to observe and its measurement precision of refractive index can meet the requirements of general scientific research and experiments.