浏览全部资源
扫码关注微信
北京理工大学 光电学院 北京,100081
纸质出版日期:2014-03-25
移动端阅览
禹璐,程德文,周伟等. 硬性内窥镜光学系统的杂散光分析与抑制[J]. 光学精密工程, 2014,22(3): 525-532
YU Lu, CHENG De-wen, ZHOU Wei etc. Stray light analysis and suppression of rigid endoscope system[J]. Editorial Office of Optics and Precision Engineering, 2014,22(3): 525-532
基于光学系统结构研究了硬性内窥镜光学系统内部杂散光的成因及抑制方法
以提高其成像效果。通过Lighttools软件进行光线追迹
分析了硬性内窥镜光机结构的杂散光;通过杂光路径分析
提出光线在中继系统和物镜组中发生全反射是硬性内窥镜存在杂散光的原因
由此研究了相应的杂散光抑制方法:即调整物镜组结构控制系统中光线传播的路径
并在光学系统优化过程中加入相应的控制条件。最后
采用提出的方法设计了视场角为70
透镜直径为2.7 mm的高成像性能消杂散光硬性关节镜。验证实验显示
研制的硬性内窥镜光学系统在100 lp/mm 处各视场的调制传递函数(MTF)值均在0.4以上
逼近衍射极限。光线追迹结果表明
在物镜组中消除杂散光可避免光线在棒镜内壁全反射
完全消除杂散光并保证良好的像质。
On the structure of optical system
the causes of stray light of a rigid endoscope and its suppression methods were explored to eliminate the stray light and to improve its imaging performance. Firstly
the stray light in the optical and mechanical system was simulated and analyzed by ray tracing in Lighttools. By analysis of the stray light path
it points out that the stray light of the optical system comes from the total reflection of a relay system and an objective system. Then
with the possible causes
the suppression methods for the stray light were proposed
including adjusting objective structure and controlling ray path. Finally
a rigid arthroscopy with a Field of View(FOV) of 70 and an outer diameter of 2.7 mm was designed to suppress the stray light. With proposed methods
the Modulation Transfer Function(MTF) reaches diffraction limit
the values of all fields are higher than 0.4 at 100 lp/mm
and the size of spot diagram is smaller than 4 m. Besides the high-definition imaging
simulation demonstrates that the stray light has been moved out entirely in the optical system
and the system shows excellent imaging quality.
李景艳, 刘德森, 刘刚, 等.医用内窥镜光学系统的应用及发展趋势[J].医疗装备, 2005, 18(7):9-12.
LI J Y, LIU D S, LIU G, et al..Application and development trend of endoscope optical system[J].Chinese Journal of Medical Device, 2005, 18(7):9-12.(in Chinese)
TOMKINSON T H, BENTLEY J L, CRAWFORD M K, et al..Rigid endoscopic relay systems:a comparative study[J].Applied Optics, 1996, 35(34):6674-6683.
PAYNE S, EARDLEY I, O′FLYNN K.Imaging and Technology in Urology:Principles and Clinical Applications[M].New York:Springer, 2012.
原育凯.光学系统杂散光的消除方法[J].大气与环境光学学报, 2007, 2(1):6-10.
YUAN Y K.Stray light suppressing measure of optical system[J].Journal of Atmospheric and Environment Optics, 2007, 2(1):6-10.(in Chinese)
岑兆丰, 李晓彤, 朱启华.光学系统杂散光分析[J].红外与激光工程, 2007, 36(3):300-304.
CEN ZH F, LI X T, ZHU Q H.Stray light analysis for optical system[J].Infrared and Laser Engineering, 2007, 36(3):300-304.(in Chinese)
DUAN J, MEI C, JIANG K, et al..The influence of the stray light on MTF in optical system in test range [C].Information Optoelectronics, Nanofabrication and Testing, Wuhan: OSA, 2012:IF4A.26.
SCHAUB M P.The Design of Plastic Optical Systems [M].New York:SPIE Press, 2009.
张军强, 吴清文, 颜昌翔.星载成像光谱仪杂散光检测技术[J].中国光学, 2010, 3(4):337-342.
ZHANG J Q, WU Q W, YAN CH X.Stray light measurement technologies for space-borne imaging spectrometer[J].Chinese Journal of Optics, 2010, 3(4):337-342.(in Chinese)
李晓平, 沙晟春, 胡亭亮.长焦距离轴三反光学系统杂散光的抑制[J].光学 精密工程, 2011, 19(11):2603-2608.
LI X P, SHA SH CH, HU T L.Suppression of stray light for long focal length off-axis three-mirror optical system [J].Opt. Precision Eng., 2011, 19(11):2603-2608.(in Chinese)
CASEMENT S, FLANNERY M, GLASSMAN T, et al..Starshade design driven by stray light from edge scatter[J].SPIE, 2012,8442:8442H-1.
李岩, 刘剑峰.红外光学遥感器内杂散光和外杂散光的综合抑制研究[J].光学学报, 2013, 33(9):928002.
LI Y, LIU J F.Research on integrative suppression of internal and external stray light in infrared optical remote sensor[J].Acta Optica Sinica, 2013, 33(9):0928002.(in Chinese)
HUANG Z, BAI J, LU T X, et al..Stray light analysis and suppression of panoramic annular lens[J].Optics Express, 2013, 21(9):10810-10820.
黄钦煊, 李步洪, 王秀琳, 等.生物组织光学性质与表面反射率的相关性[J].福建师范大学学报:自然科学版, 2000, 4(4):41-43.
HUANG Q X, LI B H, WANG X L, et al..Relationship between optical properties of tissue and their reflectance[J].Journal of Fujian Normal University:Natural Science, 2000, 4(4):41-43.(in Chinese)
张镇西, 蒋大宗, 张志麟, 等.生物组织光传播特性的研究[J].中国生物医学工程学报, 1993, 12(3):195-202.
ZHANG ZH X, JIANG D Z, ZHANG ZH L, et al..Study of optical propagation properties in biodogical tissue[J].Chinese Journal of Biomedical Engineering, 1993, 12(3):195-202.(in Chinese)
金转裕, 王海云.光学透镜的消杂光黑漆的研究[J].光学 精密工程, 1987, 2(2):43-50.
JIN ZH Y, WANG H Y.Research on black coating for restraining stray-light from optical lenses[J].Opt. Precision Eng., 1987, 2(2):43-50.(in Chinese)
LIANG R.Optical Design for Biomedical Imaging [M].New York:SPIE Press, 2010.
0
浏览量
59
下载量
1
CSCD
关联资源
相关文章
相关作者
相关机构